14.3.4.1
Reflex Threshold: Control Panel settings
Reflex Threshold
Probe Tone
Select the frequency for this measurement setup.
226 Hz
Adult default
If 226 Hz is selected, you can set admittance to be shown in the
admittance unit mmho or in the compliance volume units ml, cc or
cm
3
as described in
Reflex Threshold view settings
.
For frequencies higher than 226 Hz, admittance is shown
only
in
mmho. The value in
Y scale
automatically changes to mmho.
1000 Hz
Infant default
Recommended for testing on infants younger than 4-6 months. See
also the MADSEN OTOflex 100 Guide.
Custom TPP offset
Select or deselect the use of pressure offset to stabilize the eardrum.
To change the current setting, if
Control Panel
>
Custom TPP offset
is checked, select
Tools
>
Options
>
Reflex Threshold
>
Meas-
urement
>
General
>
Custom TPP offset
.
Disabled
: TPP is automatically set to
25 daPa more
.
Enabled
: TPP + TPP offset (according to TPP setting in
Tools
>
Options
>
Reflex Threshold
>
Measurement
>
Custom TPP off-
set
) are used.
Default value:
Enabled
Manual
Stimulus
Select the stimulus types to be included.
Available types: 500 Hz, 1000 Hz, 2000 Hz, 3000 Hz, 4000 Hz, BBN,
LBN, HBN.
Adult default: 1000 Hz
Infant default: BBN
Intensity
Select the stimulus intensity from the drop-down list.
Adult default: 70 dB HL
Infant default: 70 dB HL
•
Stimulus levels at or above the warning level
(> 108 dB SPL re 2cc) are marked by an orange bar in the
Intens-
ity
drop-down list.
•
Stimulus levels at or above the intensity safety level
(> 115 dB SPL re 2cc) are marked by a red bar in the
Intensity
drop-down list.
Otometrics - MADSEN OTOflex 100
155
14 Immittance module settings