Probe problems can be caused by:
•
an occluded probe
•
a blocked wax filter
Leakage and probe occlusion is shown on the screen during testing.
Probe status
If there is leakage or if the probe is blocked during testing, the window in the top right corner of the screen visualizes the
problem. Adjust the position of the probe, or clean the probe, and continue testing.
OTOsuite
Description
The probe seal is OK
The probe is blocked
There is a probe leak
The probe is not fully inserted
11.11
Testing the other ear
When you have finished testing one ear, select the other ear, if needed:
1. Toggle the
Ear Selection
icon on the OTOsuite toolbar before testing the other ear.
The icon shows the currently active probe ear.
2. Switch the probe (and, if needed, insert phone) to the opposite ear(s).
3. Continue testing.
11.12
Saving measurements
The Immittance module keeps 3 curves for each ear. When you have made the number of measurements you need, you
can save the curves.
1. To save the patient folder, select
File
>
Save
, or click the
Save measurements
icon on the OTOsuite
toolbar.
You can save 3 separate tympanometric measurement results for each ear. If you wish to delete any of the
curves, see
.
Note
•
Contralateral measurements, when saved, apply to the stimulus ear.
Otometrics - MADSEN OTOflex 100
115
11 Testing with the OTOsuite Immittance module