Procedure
1. Press
PageDown
to go to the Reflex Threshold test
screen, or select the test from
ViewImmittance
.
If available, the Reflex results shown in the Reflex
Threshold view are reflex screening results. If you
subsequently make a Reflex Threshold test using the
same stimulus type, the results from this test will
replace the ones shown in this view. Any stimulus
types not tested during a Reflex Threshold test will
remain in the Reflex Screening results table.
The bold horizontal line in each graph indicates the
predefined reflex criterion. It is shown in relation to
the base line marker on the vertical axis. The dis-
tance between the two is influenced by the setting
in
Tools
>
Options
>
Reflex Threshold
>
General -
Reflex criterion (mmho)
.
Warning
•
If the patient is troubled by the test, stop the test. The test is interrupted
and the pump pressure is relieved immediately.Already measured results are kept.
or
Reflex Threshold automatically follows the tympanometry sweep if you start the T+RT sequence (see
).
Automatic RT testing
1. Click
Start
in the
Auto
section of the Control Panel to start a fully automatic reflex
threshold search for all preset stimuli.
Semi-automatic RT testing
1. Click
Start
in the
Manual
section of the Control Panel to start a semi-automatic reflex
threshold search for the stimulus type selected on the Control Panel. The settings in
Tools
>
Options
>
Reflex Threshold
>
Stimulus types
>
Control the threshold search
.
Manual RT testing
If a threshold is not found during automatic or semi-automatic testing, you can test specific
stimulus intensities manually.
Otometrics - MADSEN OTOflex 100
103
11 Testing with the OTOsuite Immittance module