8.5
Tympanometry
Note
•
It is recommended that you perform a tympanometric test before making any acoustic reflex meas-
urement, and determine the acoustic reflex threshold before making a reflex decay measurement.
8.5.1
Diagnostic tympanometry
A.
Press to start the test
Note
•
It is recommended that you perform a tympanometric test before making any acoustic reflex meas-
urement, and determine the acoustic reflex threshold before making a reflex decay measurement.
The default setup available in the tympanometry test is designed for any of the tympanometry tests you can
perform. Adjust the settings to suit your purposes.
Preparing for testing
1. Do as described in
.
2. If contralateral stimulation is used in the test, make sure that both the ipsilateral probe and the contralateral phone
are in place before you start the measurement.
3. Ask the patient to sit very still and quiet during the test, without moving head or jaw.
4. If needed, toggle the
Ear Selector
to select the ear on which you wish to start the test.
Procedure
1. Select the
Tympanometry
test screen.
–
Press the right-hand softkey to start a sweep
or
–
Press and hold to start a sequence.
Warning
•
If the patient is troubled by the test, stop the test. The test is interrupted and
the pump pressure is relieved immediately.Already measured results are kept.
or
Otometrics - MADSEN OTOflex 100
67
8 Testing with MADSEN OTOflex 100