11.6
Reflex Threshold testing
Note
•
It is recommended that you perform a tympanometric test before making any acoustic reflex meas-
urement, and determine the acoustic reflex threshold before making a reflex decay measurement.
Before reflex testing
1. Do as described in
.
1. Inform the patient about the high stimulus levels in the test.
2. Ask the patient to sit very still and quiet during the test, without moving head or jaw.
3. If contralateral stimulation is used in the test, make sure that both the ipsilateral probe and the contralateral phone
are in place before you start the measurement.
4. If needed, toggle the
Ear Selector
to select the ear on which you wish to start the test.
During testing
Warning
•
If the patient is troubled by the test, stop the test. The test is interrupted and
the pump pressure is relieved immediately.Already measured results are kept.
or
11.6.1
Reflex Threshold - procedure
Note
•
To avoid automatic testing being interrupted because of high stimulus intensity levels when reaching the
warning limits, it is recommended that you set the max. intensity to 95 dB HL.
102
Otometrics - MADSEN OTOflex 100
11 Testing with the OTOsuite Immittance module