Operator’s Manual – OPTI CCA-TS
4-11
4 CALIBRATION AND QUALITY CONTROL
Fig. 4-22 Place Control
Fig. 4-20 Close Cover
Fig. 4-21 Control Calibration
10. Close the SMC cover (Fig. 4-20).
11. The system starts to calibrate (Fig. 4-21).
The green status light is now lit, indicating that
a measurement is occurring and that the sample
chamber cover should not be opened.
NOTE: If the sample measurement chamber cover
is opened while the green status light is
blinking, the cassette calibration will
be cancelled and the cassette must be
discarded.
12. Calibration is complete and it is time to place a
sample (Fig. 4-22).
•
Remove an ampoule from the box of
controls and invert gently to resuspend the
scattering particles, being careful not to heat
it with your hands.
NOTE: Do not shake ampoule vigorously.
Excessive bubble formation may affect
results.
•
Gently tap the head of the ampoule with
your fingernail to remove any liquid.
•
Carefully open the ampoule by breaking off
the top.
NOTE: Protect your fingers by using gloves or tissue
while breaking ampoule.
Summary of Contents for OPTI CCA-TS
Page 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Page 2: ......
Page 4: ......
Page 10: ......
Page 18: ......
Page 28: ......
Page 40: ......
Page 69: ......
Page 71: ......
Page 89: ......
Page 105: ......
Page 118: ......
Page 155: ......
Page 171: ......
Page 173: ......
Page 182: ......
Page 190: ......
Page 198: ......
Page 206: ......
Page 210: ......
Page 214: ......
Page 218: ......
Page 226: ......
Page 239: ......
Page 253: ......
Page 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Page 255: ......
Page 257: ......