Operator’s Manual – OPTI CCA-TS
3-17
3 CUSTOMIZATION
The OPTI CCA-TS has three types of security:
1.
A
Password Function
to limit access to various
system functions (See section 3.3.3.1.1).
2.
QC Lockouts
to help hospitals meet their QC
policies (See section 3.3.3.1.2).
3.
A
Secure Operator ID
function to limit access
to analyzer to properly trained, authorized users
(See section 3.3.3.1.3).
•
From the
<Ready>
display, select
<System Manager>
and
<Setup>
.
•
In the
<System Setup>
menu, press the
<Miscellaneous>
tab, and select
<Security>
(Fig. 3-33).
3.3.3.1.1 Setting Up a Password
The OPTI CCA-TS has a password function which,
when activated, will deny access to the setup menus
and certain database functions. The correct password
will have to be entered to grant access to these
menus and functions. The password is especially
useful to ensure that only authorized operators
can alter customized settings. The factory default
password is
404
. The factory-set password can be
changed to any number between 0 and 9999 (up to 4
digits).
1. Select
<Password Enable>
in the
<Security>
menu (Fig. 3-34). You will then
be able to enter a number (1-4 digits) in the
<Setup PW>
field.
2. Press
to enter the numbers and press
to accept the changes.
CAUTION:
Make sure the password is kept
confidential and in a safe place.
Passwords can not be retrieved!
3. Press
to return to the
<Setup>
screen
or
to return to
<Ready>
.
Fig. 3-33 Select Security
Fig. 3-34 Enter Password
3.3.3 Miscellaneous
3.3.3.1 Setting up Security
Summary of Contents for OPTI CCA-TS
Page 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Page 2: ......
Page 4: ......
Page 10: ......
Page 18: ......
Page 28: ......
Page 40: ......
Page 69: ......
Page 71: ......
Page 89: ......
Page 105: ......
Page 118: ......
Page 155: ......
Page 171: ......
Page 173: ......
Page 182: ......
Page 190: ......
Page 198: ......
Page 206: ......
Page 210: ......
Page 214: ......
Page 218: ......
Page 226: ......
Page 239: ......
Page 253: ......
Page 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Page 255: ......
Page 257: ......