Operator’s Manual – OPTI CCA-TS
3-23
3 CUSTOMIZATION
3.3.3.4 Battery Saver
Fig. 3-40 Backlight Auto-Off
This menu allows you to set options that will help
conserve power to extend battery life. The battery
saver options are only active if the analyzer is
operated from the battery.
1. From the
<Ready>
display, select
<System Manager>
and
<Setup>
.
2. In the
<System Setup>
menu, press the
<Miscellaneous>
tab and then
<Hardware>
.
3. In the
<Hardware>
screen, select the following
options (Fig. 3-40):
•
<Backlight AutoOff Enable>
- If battery
save mode is also enabled, the backlight of the
display will turn off automatically after the
selected duration.
NOTE: The screen will appear dark when the
analyzer is in standby mode and Backlight
Auto Off is enabled.
Press the touch screen to exit standby mode
and resume normal operation.
•
<Battery Save Mode>
will put the instrument
into standby mode which turns off the sample
measurement chamber heaters to conserve the
battery. Battery save mode has the following
options:
-
<Off>
- The system will always stay on in
the
<Ready>
mode until the operator turns
the system off.
-
<10 Min>
- The system will go to standby
mode after 10 minutes of inactivity.
-
<60 Min>
- The system will go to standby
mode after 60 minutes of inactivity.
4. Press
to accept the changes.
5. Press
to return to the
<Setup>
screen
or
to return to
<Ready>
.
Summary of Contents for OPTI CCA-TS
Page 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Page 2: ......
Page 4: ......
Page 10: ......
Page 18: ......
Page 28: ......
Page 40: ......
Page 69: ......
Page 71: ......
Page 89: ......
Page 105: ......
Page 118: ......
Page 155: ......
Page 171: ......
Page 173: ......
Page 182: ......
Page 190: ......
Page 198: ......
Page 206: ......
Page 210: ......
Page 214: ......
Page 218: ......
Page 226: ......
Page 239: ......
Page 253: ......
Page 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Page 255: ......
Page 257: ......