Operator’s Manual – OPTI CCA-TS
pH-1
ANALYTES
pH
pH
Clinical Significance
1
The pH value of the blood, serum or plasma may be the single most valuable factor in the evaluation of
the acid-base status of a patient. The pH value is an indicator of the balance between the buffer (blood),
renal (kidney) and respiratory (lung) systems, and one of the most tightly controlled parameters in the
body. The causes of abnormal blood pH values are generally classified as:
a) primary bicarbonate deficit - metabolic acidosis
b) primary bicarbonate excess - metabolic alkalosis
c) primary hypoventilation - respiratory acidosis
d) primary hyperventilation - respiratory alkalosis
An increase in blood, serum or plasma pH (alkalemia) may be due to increased plasma bicarbonate,
or a feature of respiratory alkalosis due to an increased elimination of CO
2
, due to hyperventilation.
A decreased pH value (acidemia) in blood, serum or plasma may occur due to an increased formation
of organic acids, an increased excretion of H
+
ions in certain renal disorders, an increased acid intake
such as in salicylate poisoning or loss of alkaline body fluids. Respiratory acidosis is the result of a
decreased alveolar ventilation and may be acute; as the result of pulmonary edema, airway obstruction or
medication, or may be chronic; as the result of obstructive or restrictive respiratory diseases.
Measurement Principle
The pH optode measurement principle
is based upon pH-dependent changes of the luminescence of a
dye molecule immobilized in the optode. Such pH indicator dyes have been used by chemists for many
years to perform acid-base titration in turbid media.
The relationship of luminescence to pH is quantified by a variant of the Mass-Action Law of chemistry,
I
0
/ I = 1 + 10
pKa-pH
which describes how the fluorescence emission intensity increases as the blood pH is increased above the
dye’s characteristic pKa
2.
pH optodes do not need a reference electrode to measure pH, however, they
exhibit a small sensitivity to the ionic strength of the sample being measured
3.
Measurement Range
Range
Resolution (Low/High)
Units
6.6 to 7.8
0.01/0.001
pH units
Standard Reference Cassette (SRC) Limit Values
LOW
NORMAL
HIGH
Units
7.100 ± 0.02
7.400 ± 0.02
7.600 ± 0.02 pH units
Summary of Contents for OPTI CCA-TS
Page 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Page 2: ......
Page 4: ......
Page 10: ......
Page 18: ......
Page 28: ......
Page 40: ......
Page 69: ......
Page 71: ......
Page 89: ......
Page 105: ......
Page 118: ......
Page 155: ......
Page 171: ......
Page 173: ......
Page 182: ......
Page 190: ......
Page 198: ......
Page 206: ......
Page 210: ......
Page 214: ......
Page 218: ......
Page 226: ......
Page 239: ......
Page 253: ......
Page 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Page 255: ......
Page 257: ......