Operator’s Manual – OPTI CCA-TS
7 DIAGNOSTICS AND TROUBLESHOOTING
7.3 Troubleshooting Procedure for tHb/SO
2
If your OPTI fails an Hb calibration or QC measurement for tHb or
S
O
2
, OPTI
Medical recommends that you clean the SMC cover of your analyzer and then
repeat the measurement. The two small optical channels pictured below are re-
sponsible for the tHb and
S
O
2
measurements. These channels may get clogged or
dirty, causing the tHb and
S
O
2
to fail calibration or OPTI Check controls.
The simple cleaning procedure below can be used for OPTI CCA, OPTI CCA-
TS, or OPTI R analyzers and may correct tHb and
S
O
2
failures.
1. Open the SMC cover and locate the two small optical channels pictured
below (Fig. 7-58).
Fig. 7-58 Locate Optical Channels
2. Clean the optical channels using a cotton swab or lint-free cloth dipped in
alcohol (Fig. 7-59).
Fig. 7-59 Clean Optical Channels
Please contact OPTI Medical at +1 770.510.4444, or 800.490.OPTI (6784) or
[email protected] for any additional questions or information
regarding this procedure.
Summary of Contents for OPTI CCA-TS
Page 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Page 2: ......
Page 4: ......
Page 10: ......
Page 18: ......
Page 28: ......
Page 40: ......
Page 69: ......
Page 71: ......
Page 89: ......
Page 105: ......
Page 118: ......
Page 155: ......
Page 171: ......
Page 173: ......
Page 182: ......
Page 190: ......
Page 198: ......
Page 206: ......
Page 210: ......
Page 214: ......
Page 218: ......
Page 226: ......
Page 239: ......
Page 253: ......
Page 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Page 255: ......
Page 257: ......