
Operator’s Manual – OPTI CCA-TS
7-17
7 DIAGNOSTICS AND TROUBLESHOOTING
7.2.9 Checking the Valve Drive
Fig. 7-10 Valve Drive
7.2.8 Checking the Gas Valve
The purpose of this test is to check for proper
function of the gas valve.
From the
<Ready>
display, press
<System
Manager>
and
<Diagnostics>
.
•
Select the
<Control>
tab.
•
Press the
<Open/Closed>
button under
<Gas Valve>
to start the test (Fig. 7-9).
•
A faint hissing sound may be heard with the
pump cartridge removed and the gas valve open.
•
Press
to return to the
<System>
screen.
Fig. 7-9 Gas Valve
This diagnostic checks the proper operation of the
cassette valve drive mechanism.
From the
<Ready>
display, press
<System
Manager>
and
<Diagnostics>
.
•
Select the
<Control>
tab.
•
Press
to start the test (Fig. 7-10).
•
Enter the various positions (allowed positions
are 1 - 6) and verify the valve drive moves
smoothly and precisely.
•
Press
to return to the
<System>
screen.
Summary of Contents for OPTI CCA-TS
Page 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Page 2: ......
Page 4: ......
Page 10: ......
Page 18: ......
Page 28: ......
Page 40: ......
Page 69: ......
Page 71: ......
Page 89: ......
Page 105: ......
Page 118: ......
Page 155: ......
Page 171: ......
Page 173: ......
Page 182: ......
Page 190: ......
Page 198: ......
Page 206: ......
Page 210: ......
Page 214: ......
Page 218: ......
Page 226: ......
Page 239: ......
Page 253: ......
Page 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Page 255: ......
Page 257: ......