Na-2
Operator’s Manual – OPTI CCA-TS
SODIUM
ANALYTES
Interferences
The OPTI CCA-TS Na
+
sensor has no measurable interference from K+ variation within the range 0.8-10
mmol/L.
The OPTI CCA-TS Na
+
sensor does exhibit a small interference from Li+. Li+ levels of 1.0, 2.5, and
6.4 mmol/L will cause a positive Na+ bias of 0.9, 1.2, and 1.3 mmol/L, respectively. A syringe sample
anticoagulated with typical amounts of lithium heparin has 1-4 mmol/L of lithium, which offsets the
measured Na+ by less than 1%.
To minimize the interference from lithium, use syringes containing the lowest acceptable heparin level.
Carefully follow the syringe manufacturer’s recommendation regarding proper filling of the syringe.
A partially filled syringe results in excessive lithium concentration.
The OPTI CCA-TS Na
+
results include an appropriate correction for pH at all values of pH.
This correction may introduce an extra source of variability at the extreme values.
Heparin salts are the only acceptable anticoagulants. Other anticoagulants such as citrate, EDTA, oxalate,
and fluoride cause significant interferences to the electrolyte sensors.
The following exogeneous interferents were quantified in tonometered plasma, showing interferences to
dyes which typically have short half-lives within the body before being metabolized by the liver.
Substance
amount
Na
+
change (mmol/L)
Sodium fluorescein
26 mg/dL
unstable
Cardio (indocyanine) green
0.5 mg/dL
-18
Methylene blue
25 mg/dL
-2
Only clear, uncolored quality control materials, such as OPTI CHECK or OPTI CHECK PLUS brand
aqueous controls should be used with the OPTI CCA-TS system. Colored materials, including proficiency
testing materials, may interfere with the ion measurement, or fail to be properly aspirated.
Reproducibility
Typical Within-Run (Swr), Between-Day (Sdd) and Total (ST) Precision is determined from 1 run per day
with 2 replicates per run for 20 days on each of two OPTI CCA instruments. Sodium values are expressed
in mmol/L.
Material
mean S
wr (CV%)
S
dd (CV%)
S
T (CV%)
OPTI CHECK PLUS
(aqueous control solution)
Level 1
117.3
0.10 (0.1)
0.49 (0.4)
0.50 (0.4)
Level 2
142.3
0.11 (0.1)
0.50 (0.4)
0.51 (0.4)
Level 3
158.6
0.11 (0.1)
0.70 (0.4)
0.72 (0.5)
Serum
151.9
0.08 (0.1)
0.92 (0.6)
0.92 (0.6)
Reduced Bovine Hemoglobin
122.3
0.09 (0.1)
0.64 (0.5)
0.65 (0.5)
Solution
Summary of Contents for OPTI CCA-TS
Page 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Page 2: ......
Page 4: ......
Page 10: ......
Page 18: ......
Page 28: ......
Page 40: ......
Page 69: ......
Page 71: ......
Page 89: ......
Page 105: ......
Page 118: ......
Page 155: ......
Page 171: ......
Page 173: ......
Page 182: ......
Page 190: ......
Page 198: ......
Page 206: ......
Page 210: ......
Page 214: ......
Page 218: ......
Page 226: ......
Page 239: ......
Page 253: ......
Page 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Page 255: ......
Page 257: ......