ANALYTES
Correlation to Other Methods
4
OPTI CCA-TS vs other PO
2
Instruments on whole blood in a typical setting
PO
2
analysis of heparinized whole blood samples was performed at multiple clinical sites. Samples were
analyzed on the OPTI CCA-TS in parallel with laboratory instrumentation operated by hospital personnel
and controlled following the hospital’s established procedures.
Comparative Method*
Slope
Intercept
R
2
Sy.x
Range
n
OPTI R
1.04
-2.76
0.968
8.72
27 - 288
148
Analyzer A
0.97
3.73
0.992
5.24
27.0 – 423.8
110
* For more information on specific analyzers used, please contact OPTI Medical Technical Support.
OPTI CCA-TS with B-Lac Cassette vs other
P
O
2
Instruments on whole blood
(in-house testing)
Whole blood samples from multiple donors were tonometered with different %O
2
gas mixtures to
generate a wide range of
P
O
2
values. The blood samples were analyzed in parallel on the B-Lac cassette
and other laboratory instruments.
Comparative Method*
Slope
Intercept
R
2
Sy.x
Range
n
OPTI CCA (Std.
P
O
2
sensor) 0.94
4.84
0.998
9.77
19.9 - 642.8
161
Analyzer B
0.95
6.32
0.992
18.25
17.0 - 635.7
161
* For more information on specific analyzers used, please contact OPTI Medical Technical Support.
References
1. Tietz, Norbert W., Ed.,
Clinical Guide to Laboratory Tests
, 2nd Ed., (Philadelphia: W.B.Saunders,
Co., 1990) p. 436
.
2. Kautsky H,
Quenching of Luminescence by Oxygen
, Transactions Faraday Society 35, p.216, 1939
3. CDI,
3M Healthcare System 200 Extracorporeal Blood Gas Monitor
. See, for example, Lubbers DW,
Gehrich J, Opitz N,
Fiber Optics Coupled Flourescence Sensors for Continuous Monitoring of Blood
Gases in the Extracorporeal Circuit,
Life Supports Systems 4, p.94, 1986.
4. OPTI Medical.
Model equation for regression statistics is: [results of OPTI Analyzer] = slope(m)
[comparative method results] + intercept(b)
.
Summary of Contents for OPTI CCA-TS
Page 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Page 2: ......
Page 4: ......
Page 10: ......
Page 18: ......
Page 28: ......
Page 40: ......
Page 69: ......
Page 71: ......
Page 89: ......
Page 105: ......
Page 118: ......
Page 155: ......
Page 171: ......
Page 173: ......
Page 182: ......
Page 190: ......
Page 198: ......
Page 206: ......
Page 210: ......
Page 214: ......
Page 218: ......
Page 226: ......
Page 239: ......
Page 253: ......
Page 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Page 255: ......
Page 257: ......