Operator’s Manual – OPTI CCA-TS
pH-B-3
ANALYTES
pH (Dry Sensor)
The following samples were identified as interfering with the dry pH sensor in the interference study
performed for the OPTI LION 510(k) submission.
Interferent
Test Level
Change
Sodium Bisulphate
11.5mM
-0.16
Phenylacetic Acid
10.0mM
-0.12
Methylene Blue
25mg/dL
Unstable
Fluorescein
25mg/dL
Unstable
Reproducibility
Controls
Within-Run (Swr) and Total (ST) Precision were determined from 2 runs per day with 2 replicates per run
over a period of 20 days following the CLSI guideline EP5-A2. Typical results for three different lots of
B-Lac cassettes are shown below.
Material
mean
Lot 1
Lot 2
Lot 3
OPTI CHECK
(aqueous control solution)
Swr
ST
Swr
ST
Swr
ST
Level 1
7.17
0.011
0.014
0.008
0.007
0.005
0.006
Level 2
7.44
0.010
0.010
0.010
0.010
0.004
0.005
Level 3
7.65
0.011
0.011
0.007
0.007
0.004
0.006
Whole Blood
Within-Run precision in whole blood samples was evaluated at three different pH concentrations using
multiple instruments and multiple cassette lots.
pH in Whole Blood
Level 1
Level 2
Level 3
Average
7.170
7.306
7.610
St. Dev
0.008
0.007
0.011
%CV
0.12%
0.10%
0.14%
n
18
18
18
Linearity
3
Wherever possible, linearity for the OPTI CCA-TS measurement has been established against reference
materials or methods. The linearity of the dry pH sensor has been established versus the standard
pH sensor which is already approved for use on the OPTI CCA-TS system. Whole blood samples
tonometered with different %CO
2
gas mixtures were used to establish the correlation.
Linearity of Whole Blood Samples
Slope
Intercept
R
2
Sy.x
Range
n
0.99
0.09
0.996
0.016
6.59-7.86
189
Summary of Contents for OPTI CCA-TS
Page 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Page 2: ......
Page 4: ......
Page 10: ......
Page 18: ......
Page 28: ......
Page 40: ......
Page 69: ......
Page 71: ......
Page 89: ......
Page 105: ......
Page 118: ......
Page 155: ......
Page 171: ......
Page 173: ......
Page 182: ......
Page 190: ......
Page 198: ......
Page 206: ......
Page 210: ......
Page 214: ......
Page 218: ......
Page 226: ......
Page 239: ......
Page 253: ......
Page 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Page 255: ......
Page 257: ......