Getting Started FLUOVIEW/
Outline of LSM Observation Procedures
OPERATION INSTRUCTIONS
1 - 2 7
Page
One Point!
After the dyeing method has been set with the [Dyes] sub-panel, the ND filters can be
set using the [Ch] group box in the upper part of the [Acquire] panel.
1.
In the upper part of the [Acquire] sub-panel, open the [Ch] group box for the ND to be
changed.
2.
Click the <More> button.
The field as shown below is displayed below the [Ch] group box.
3.
Vary the ND value using the Laser LED slider.
Each click of the laser ND <+> or <-> varies the laser ND value by 1%.
Each click of the laser ND slider varies the laser ND value by 5%.
1-2-7 Setting the Observation Condition
1-2-7-1 Setting the Objective Magnification
From the drop-down list on the center of the [Acquire] panel, select the objective being used
with the microscope.
If you change the objective, click the <Apply> button in the [Dyes] sub panel.
The Conforcal Aperture value is set appropriately.
For detailes, see section 1-3-2-4 “Configuring the Filters” for automatic
Confocal Aperture setting with switching the objective.
Clicking this button allows fine
adjustment of the ND value.
Clicking this field allows the
ND value to be changed on a
large scale.
The ND values which are
usually used are displayed in
green.
Display of the ND value set by
clicking the <+> and <-> buttons or
the field. The set ND value can
also be changed by entering its
value directly from the keyboard.
Display of the optimum laser
and ND value which are set
automatically according to the
selected dyeing method.