APPLIED OPERATIONS
/ Protocol processor
OPERATION INSTRUCTIONS
2 - 1 9 2
Page
z
Setting Track 3
Acquire nine images at one-hour interval in each Track in XYT observation.
1. Select “New mode”-“XYT” in [Mode] of [Track].
2. Enter “Wait 3600” in [Stand by] of [Track]. (Enter the value by the second.)
3. Enter “9” in [Condition] of [Track].
4. Select “Middle” in [Scan Speed].
5. Enter the laser intensity value for use in image acquisition in the cell for each laser in
[Laser Intensity].
Enter “8” in [Ar] and “50” in [HeNe-G].
6. Enter the PMT value of each channel in the cell for each channel in [PMT Voltage].
Enter “Init” in [PMT2] and [PMT3].
7. Enter the image file name to be stored to [Save FileName.].
z
Setting Track 4
Specifies the repeated protocol by use N variable.
1. Select “Next” in [Mode] of [Track].
2. It is not necessary to enter the values in other cells.