APPLIED OPERATIONS
/Image Processing
OPERATION INSTRUCTIONS
2 - 2 3 6
Page
6.
Enclose the histogram section of interest using the scale in the [Master View] field.
7.
The magnified view of the region selected by the scale is shown in the [Selected View]
field.
Fig. 2-89 Histogram of the Selected Region
8. When the image was acquired in the multi-channel mode, select the channel(s) to be
subjected to LUT intensity mapping (re-assignment) using the [Channel] option buttons.
9. Select the region to be subjected to mapping (re-assignment) using the scale in the
[Selected View] field. While moving the scale, confirm the change in contrast in the
image window. The intensity values in the selected region are mapped (re-assigned) to
intensity values from 0 to 255 and displayed.
The vertical chart axis
represents the degree and
the horizontal axis
represents the mapped
intensity values.