APPLIED OPERATIONS
/Image Acquisition
OPERATION INSTRUCTIONS
2 - 8 7
Page
8. Click the <XY Repeat> or <Focus> button to perform repeated scanning in order to
adjust the brightness of the channels.
Select a group before repeated scanning. The group selection cannot be
changed during repeated scanning.
9.
Adjust the brightness using the [PMT], [Offset] and [Gain] LED sliders in the [Ch] group
box in the [Acquire] panel. For details, see section 2-2-1-4-9, “Adjusting the image
brightness”.
10. Click the <STOP SCAN> button to stop repeated scanning.
11. Repeat steps 7 and 10 for each of the groups to be configured.
12. Click the <Seq. Once> button to start sequential scan.
13. After image acquisition, click <Normal> below the [Surface XY] option button in the
[Mode] group box in the [Acquire] panel to return from the image acquisition mode.
This hides the virtual channels from the display.
Only the image(s) of the current Sequential Group is updated when image is
acquired by pressing <XY Repeat> button or <Focus>. The other Sequential
Group image stays in [Live] panel.
<XY Repeat> button
<Focus> button
<Seq Once> button
TIP
TIP
TIP