Operation Manual
Troubles and Countermeasures
/ When Error Codes Are Displayed
5-3
Page
5-2 When Error Codes Are Displayed
When an error occurs, the loader detects the error and stops all
operations. At the same time, a blinking error code appears on the liquid
crystal panel
and a buzzer sounds.
Press the [M1] menu button
to silence the buzzer.
When there are no wafers in the transfer path, turn the main switch OFF
and then turn it back to ON.
When there are wafers in the transfer path, turn the main switch OFF,
remove the wafer(s) using tweezers or similar tools, and ensure that the
cause is dealt with.
When an adsorption error has occurred, the loader may be unable to
hold wafers securely. Take care not to perform automatic unloading
without dealing with the cause. Wafers may be damaged.
If any similar errors occur during operation, please contact the OLYMPUS
distributor for your loader.
Please give the distributor the following information:
Product name:
Serial No.:
Error code:
Phenomenon:
E0101
OFF
Caution
Summary of Contents for AL120-6 Series
Page 2: ...This page intentionally left blank ...
Page 9: ...Operation Maintenance Manual Introduction Safety Precautions i 7 Page AL120 86 Series Unit mm ...
Page 16: ...Introduction Conformity Standards Operation Maintenance Manual i 14 Page ...
Page 17: ...Operation Maintenance Manual Introduction Conformity Standards i 15 Page ...
Page 21: ...Operation Manual Table of Contents C 3 Page 4 When a Wafer is on the Vacuum Stage 5 10 ...
Page 22: ...Operation Manual Table of Contents C 4 This page intentionally left blank Page ...
Page 58: ...Inspections Operation Manual 3 14 This page intentionally left blank Page ...