Inspections
Operation Manual
3-10
3-2-7
Top Macro Insp Back Macro Inspection
Page
h
*
This function is available for the LMB model only.
1.
Set a cassette on the cassette table.
2.
Select the [Top Macro] and [Back Macro] Observation buttons. ..................................................................
c
When the buttons are selected, the LEDs in the buttons light up.
If the [Top Macro] and [Back Macro] LEDs are not lit, press the buttons to turn on the LEDs. If LEDs for
other types of inspection are lit, turn them off by pressing their buttons.
3.
Confirm the detailed settings for inspection.
1) [Inspection Time]: Enables the setting of Macro inspection time. ...................................................................
d
* The inspection time is common to the Top Macro, Back Macro and 2nd Back Macro inspections.
2) [Top Macro spin direction]: Enables the setting of wafer rotation direction during the Top Macro inspection.
e
3) [Top Macro spin speed]: Enables the setting of wafer rotation speed during the Top Macro inspection. ......
e
4.
Set and confirm the inspection mode.
Select All or Sampling (P1 to P10). ..............................................................................................................
f
Set the inspection wafer number(s) as needed. ..........................................................................................
g
5.
Press the [Start] button to transfer the first wafer is transferred to the Top Macro inspection position. .......
h
6.
After the specified inspection time has elapsed, the macro table is lowered and the loader proceeds to
Back Macro inspection.
You can change the wafer tilt position (angle) for observation using the [Back Macro Tilt Angle] button. ...
i
* Press the [Start] button if the [Inspection Time] control is set to [
∞
].
7.
After the specified inspection time has elapsed, the wafer is unloaded into the cassette, and the second
wafer is transferred to the Top Macro inspection position.
Subsequent wafers are transferred automatically and sequentially.
8.
After the last wafer is unloaded into the cassette, the operating units return to their initial positions and the
loader stops.
c
f
g
d
i
e
Summary of Contents for AL120-6 Series
Page 2: ...This page intentionally left blank ...
Page 9: ...Operation Maintenance Manual Introduction Safety Precautions i 7 Page AL120 86 Series Unit mm ...
Page 16: ...Introduction Conformity Standards Operation Maintenance Manual i 14 Page ...
Page 17: ...Operation Maintenance Manual Introduction Conformity Standards i 15 Page ...
Page 21: ...Operation Manual Table of Contents C 3 Page 4 When a Wafer is on the Vacuum Stage 5 10 ...
Page 22: ...Operation Manual Table of Contents C 4 This page intentionally left blank Page ...
Page 58: ...Inspections Operation Manual 3 14 This page intentionally left blank Page ...