5-12. CAL-DEFAULT
▌This function is to restore to the state of calibration data that shipped from factory.
▌If you calibrate the unit in the condition that you have no previous knowledge of calibration or you
can not restore by using "5-10. CAL-RESTORE" function, you can return to factory default calibration d
If you use this function, you must calibrate this unit at the official calibration center.
CAL-DEFAULT Operating
▌Power switch on in a pressing "FACTORY" key.
▌Power switch ON
Release
▌When appearing "1. user-mem clear" message, release "FACTORY" key
▌Enter to "10. CAL-DEFAULT" mode by using cursor key.
▌In order to restore, press the "FACTORY" key one more time.
LCD Display
≫ There's no related Remote Interface Command
5-13. LOAD DEFAULD
▌This function is to return to default value that shipped the unit from factory. You can erase
setting and changing all value at the "Factory Mode"
▌Initializing contents
> User Memory Clear
> Last State : Disable
> Auto Cursor Move : Disable
> Auto Key Lock : Disable > OVP Use : Enable
> OCP Use : Enable
> ADC Sampling : 20Hz
LOAD DEFAULT Operating
▌Power switch on in a pressing "FACTORY" key
▌Power switch ON
Releasing
▌When appearing "1. user-mem clear" message, release "FACTORY" key.
▌Enter to "11. LOAD DEFAULT" mode by using cursor key.
▌In order to return to default value, press the "FACTORY" key one more time
LCD Display
≫ Related Remote Interface Command
FACT:LOAD-DEF
Apply: FACT:LOAD-DEF
Initialize all the Factory setting value
Hold press
Hold Press
- 66 -
done
done
Summary of Contents for MX Series