3-5. Programming Over Current Protection(OCP)
This function protects the DUT from power supply if the output voltage is higher than the OCP leve
The procedure below describes how to set the OCP level and turning it on and off.
The OCP level values shown below may be different from the actual values.
▌Turn on the power switch.
After turning on power, check if the unit is in "**OUTPUT OFF**" mode.
▌Press the protection key to set the levels.
Displayed LCD Message
PROTECT> 1.OVP
▌Use the right cursor key to go to the OCP menu.
Displayed LCD Message
PROTECT> 2.OCP
▌Press the protection key once more in “
protect> 2.ocp
.”
Displayed LCD Message
OcP SET> 88.00a
▌Use the cursor keys to place the cursor on the desired value.
▌Control the encoder switch to change the level value.
The OCP level cannot be set lower than the currently outputted current
to prevent operator mistakes.
▌Press the protection key when the change is complete.
Displayed LCD Message
CHANGED
Return to the previous stat after the message above is displayed.
Canceling
▌If you want to cancel the setting, wait for 5 seconds.
Displayed LCD Message
CANCEL…
≫ Related remote interface commands
CURR:OCP {<current>}
CURR:OCP?
CURR:OCP:TRIP?
CURR:OCP:CLE
Application: CURR:OCP 50
Set OCP Level
Refer to “Initial Setting Value” in “1-6 Check after Power On” for the initial OVP settings
Note
- 27 -
Wait for 5 Seconds