3-13. Store
You can save the current power supply’s state in the 『User Memory』. The 『User Memory』
is composed of 10 slots, and the set Output ON/OFF state, voltage, current, OVP, OCP, UVL,
OVL, UCL, and OCL level values are saved.
The procedure for saving the settings on the 『User Memory』 is shown below.
The Store key also provides a software calibration feature that is required for calibrating the unit.
Refer to “4. CALIBRATION” for the calibration feature
▌Turn on the power switch.
After turning on power, check if the unit is in "**OUTPUT OFF**" mode.
▌Press the STORE key to save the current power supply’s state
in the 『User Memory』.
Displayed LCD Message
STORE NO, 01
▌Use the encoder switch to select the slot you want to use among 01 ~ 10.
For example, turn the encoder switch once clockwise to save the state
in slot 02.
Displayed LCD Message
STORE NO, 02
▌Press the STORE key once more to save in the selected slot.
Displayed LCD Message
Saved
▌The LCD returns to the previous screen after displaying the "Saved” message.
≫ Related remote interface commands
*SAV {1|2|3|4|5|6|7|8|10}
Application *SAV 4
This saves the current state in the 4th 『User Memory』 slot
Saving in Output ON state can be dangerous since the output can be activated immediately during recall,
so save in OFF state for safety.
The 『User Memory』 can be reset. "Refer to 5-3. USER-MEM CLEAR"