3-14. Recall
This feature applies the settings saved in the 『User Memory』 by the STORE feature
to the current power supply. There are 10 memory slots, with the set Output ON/OFF state,
voltage, current, OVP, OCP, UVL, OVL, UCL, and OCL values applied.
The procedure for activating recall is shown below.
The unit supports a FACTORY feature as well as the RECALL feature.
Refer to "5.FACTORY “ for the factory feature
▌Turn on the power switch.
After turning on power, check if the unit is in "**OUTPUT OFF**" mode.
▌Press the RECALL key to apply the data saved in the 『User Memory』
to the current power supply.
Displayed LCD Message
RECALL NO, 01
▌Use the encoder switch to select the slot you want to use among 01 ~ 10.
For example, turn the encoder switch once clockwise to save the state
in slot 02.
Displayed LCD Message
RECALL NO, 02
▌Press the RECALL key once more to apply the settings saved in the selected slot.
Displayed LCD Message
saved
▌The LCD returns to the previous screen after displaying the "
saved
” message.
≫ Related remote interface commands
*RCL {1|2|3|4|5|6|7|8|10}
Application *RCL 4
This applies the settings saved in the 4th 『User Memory』 slot to the current state
Storing in Output ON state can be dangerous since the output can be activated immediately during
recall, For safety, store in OFF state even if it is inconvenient.
Note
- 45 -