3-8. Programming Under Current Limit (UCL)
The encoder switch can change the current from 0A up to the maximum value. If the UCL is active,
however, the encoder can change the current only from the UCL up to the maximum voltage.
The UCL level values shown below may be different from the actual values.
▌Turn on the power switch.
After turning on power, check if the unit is in "**OUTPUT OFF**" mode.
▌Since the current limit must be set higher than the set UCL value before setting
the UCL, the current limit must be set first.
▌When setting the current, use the left cursor key to move the cursor to the
desired current range and use the encoder switch to select your desired current.
The current cannot exceed the protection level if the Protection(OCP/UCL/OCL)
settings are active. If you try to exceed the limit in this case, the “PROTECTION
LIMIT” message will be displayed, and the current will be set to the protection level.
▌Press the protection key to set the UCL level.
Displayed LCD Message
PROTECT> 1.OVP
▌Use the right cursor key to go to the UCL menu.
Displayed LCD Message
PROTECT> 5.uCl
▌Press the protection key once more in “protect> 5.UCL.”
Displayed LCD Message
UCL SET> 00.00A
▌Use the cursor keys to place the cursor on the desired value.
▌Control the encoder switch to change the level value.
The UCL level cannot be set higher than the currently outputted voltage
to prevent operator mistakes.
▌Press the protection key when the change is complete.
Displayed LCD Message
CHANGED
Return to the previous stat after the message above is displayed.
Canceling
▌If you want to cancel the setting, wait for 5 seconds.
Wait for 5 Seconds
Displayed LCD Message
CANCEL…
≫ Related remote interface commands
CURR:UCL {<current>}
CURR:UCL?
Application: CURR:UCL 5
Set UCL Level
Refer to “Initial Setting Value” in “1-6 Check after Power On” for the initial OVP settings
Note
- 31 -