3-8. Programming Under Current Limit (UCL)

The encoder switch can change the current from 0A up to the maximum value. If the UCL is active, 
however, the encoder can change the current only from the UCL up to the maximum voltage.

The UCL level values shown below may be different from the actual values.

▌Turn on the power switch.

   After turning on power, check if the unit is in "**OUTPUT OFF**" mode.

▌Since the current limit must be set higher than the set UCL value before setting

 the UCL, the current limit must be set first.

▌When setting the current, use the left cursor key to move the cursor to the 

   desired current range and use the encoder switch to select your desired current.

   The current cannot exceed the protection level if the Protection(OCP/UCL/OCL) 

   settings are active. If you try to exceed the limit in this case, the “PROTECTION 
   LIMIT”  message will be displayed, and the current will be set to the protection level.

▌Press the protection key to set the UCL level.

   Displayed LCD Message

PROTECT> 1.OVP

▌Use the right cursor key to go to the UCL menu.

   Displayed LCD Message

PROTECT> 5.uCl

▌Press the protection key once more in “protect> 5.UCL.”

   Displayed LCD Message

UCL SET> 00.00A

▌Use the cursor keys to place the cursor on the desired value.

▌Control the encoder switch to change the level value.

  The UCL level cannot be set higher than the currently outputted voltage
  to prevent operator mistakes.

▌Press the protection key when the change is complete.

   Displayed LCD Message

CHANGED

   Return to the previous stat after the message above is displayed.

Canceling

▌If you want to cancel the setting, wait for 5 seconds.

Wait for 5 Seconds

   Displayed LCD Message

CANCEL…

≫ Related remote interface commands

CURR:UCL {<current>}
CURR:UCL?

Application: CURR:UCL 5

Set UCL Level

Refer to “Initial Setting Value” in “1-6 Check after Power On” for the initial OVP settings

 Note

 -  31  - 

Summary of Contents for MX Series

Page 1: ...www odacore com Switching Type Programmable DC Power Supply MX Series User Manual Manual Part NO 018MX 1 0...

Page 2: ...ontained herein Should ODA and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms the warranty terms in the separate a...

Page 3: ...of parts remove instrument covers Always disconnect Do not place power cord equipment inlet the power cable and any external circuits before main switch etc to make it difficult to operate removing t...

Page 4: ...assis terminal Terminal is at earth potential Out position of a bi stable push switch Caution risk of electric shock Standby supply Unit is not completely disconnected from ac mains when switch is off...

Page 5: ...ions 8 1 5 Check Before Power On 9 Power Cord Check Rack Mounting 1 6 Check After Power On 10 Display Procedure on the LCD Default Setting Values 1 7 Installation 11 Cooling Bench Operation 2 Front Pa...

Page 6: ...rrent Limit OCL 34 35 36 RS232C Setting RS232C Configuration RS485 Setting RS232C RS485 Connection Drawing TCP IP Setting Option Delimiter Response 45 3 13 Store 46 3 14 Recall 47 3 15 Output ON OFF 4...

Page 7: ...ory Clear Operating 5 4 LAST STATE 61 LAST STATE Operating 5 5 AUTO CURSOR MV 62 AUTO CURSOR MV Operating 5 6 AUTO KEY LOCK 63 AUTO KEY LOCK Operating 5 7 OVP USE 64 OVP USE Operating 5 8 OCP USE 65 O...

Page 8: ...ting Oprating Commands 71 6 5 Measure Commands 76 6 6 Calibration Commands 77 77 80 7 Error Messages 83 7 1 Operating Error 83 7 2 Hardware Error 83 83 85 7 5 Non volatile Memory Check Error 85 7 6 In...

Page 9: ...asy Interface Settings by Using the Front Control Panel s I O config Technical Realization of Products Insulation Floating Logic Built In Error Checking Function for SCPI Programming Grammar Calibrati...

Page 10: ...ication Manual Included Available for Download at the Homepage http www odacore com Options RS232C Cable 1M 2M 4M RS485 Cable 2 4 8 Channel 1M 2M 4M 10M TCP IP Analog Input 0 10V or 0 5V Analog Output...

Page 11: ...o and model name during the initial self testing phase If the OUTPUT OFF message is displayed without any problems refer to the specifications while checking the unit Service Center 82 32 623 5454 Hom...

Page 12: ...caused by overloading it has built in fast fuses If the unit does not turn on even after turning on power switch you can request A S from a nearby service center or the head office The AC input cable...

Page 13: ...If the Last State Mode is Disabled OVP OVP is set for the maximum value Remote Interface Set to maintain the previous settings OVP is set to ON OCP OCP is set for the maximum value State after self te...

Page 14: ...ow in the front and rear of this unit The bottom has no air vent so you can operate the unit without taking this into consideration so this must be taken into consideration when installing the unit If...

Page 15: ...e Reset or Factory Key Output Voltage Current ON OFF Key IO Local Settings and Error Display Key Protection Settings and Front panel Lock Key Voltage Current Settings Value Display Key Save Current St...

Page 16: ...e front panel cannot be used 7 Key Lock Lamp The front panel is locked and cannot be used if this lamp is on 8 Error Lamp LED is lit when various errors occur and turned off when all errors have been...

Page 17: ...and Front panel Lock Key Can set various protection functions such as OVP OCP UVL OVL UCL OCL Pressing the key for 1 second turns on the LOCK Lamp and locks the front panel Pressing the key once more...

Page 18: ...r changes the current settings 5 Rotate the voltage or current encoder clockwise to increase the limit value To decrease the limit value rotate the encoders counter clockwise 6 Check if the set values...

Page 19: ...Current Control Terminal Block Terminal Block for Controlling the Current in Proportion to Input of 0 5V or 0 10V 21 Inputs T Phase Control Port for Operating 2 or More MX Series Units Option Port s G...

Page 20: ...t has a built in RS232C RS485 connection and the OUT port is for parallel RS485 connection 4 RS485 Termination Switch On Off switch for using terminating resistors when using RS485 connection Turns on...

Page 21: ...f checking mode switch the unit to OUTPUT OFF mode and leave power supply on standby 3 Press the Output On Off Key so that the voltage is outputted from the output port 4 Press the Limit Display Key t...

Page 22: ...9 Programming Over Current Limit OCL Description for the over current limit 10 Key Lock Description on how to enable and disable the front panel lock 11 IO Local 12 Limit Display Description on viewi...

Page 23: ...Current 1 The left cursor key increases the voltage range and the right cursor key decreases the current rang 2 What is ReadBack Display This means that the LCD is displaying the currently outputted v...

Page 24: ...k if the voltage provided for the DUT is enough and increase the voltage limit value Related remote interface commands APPLy voltage current VOLT voltage UP DOWN CURR current UP DOWN OUTP OFF ON Appli...

Page 25: ...the DUT during V sensing you must add 5mV each to the 1V drop between the S point and output terminal Since the sensing cable is part of the power supply s feedback system the sensing cable s own res...

Page 26: ...splayed voltage and current values are significantly different from the output terminal s voltage and current this is very dangerous Since V sensing under the circumstances wherein the DUT s cable is...

Page 27: ...D Message CHANGED Return to the previous stat after the message above is displayed Canceling If you want to cancel the setting wait for 5 seconds Displayed LCD Message CANCEL Related remote interface...

Page 28: ...current capacity as in the figure below can prevent this problem 4 If it is difficult to resolve the trip you can turn off the OVP feature The OVP TRIPPED message will be displayed if the actual outp...

Page 29: ...nge is complete Displayed LCD Message CHANGED After the message above is displayed you will be returned to the state before the trip occurred Related remote interface commands VOLT OVP voltage VOLT OV...

Page 30: ...on key once more in protect 2 ocp Displayed LCD Message OcP SET 88 00a Use the cursor keys to place the cursor on the desired value Control the encoder switch to change the level value The OCP level c...

Page 31: ...eature switch to factory mode and select Disable in 5 OVP USE The LCD will display OCP TRIPPED if an OCP trip occurs To clear the OCP trip start with increasing the OCP level Press the protection key...

Page 32: ...TECTION LIMIT message will be displayed and the voltage will be set to the protection level Press the protection key to set the UVL level Displayed LCD Message PROTECT 1 OVP Use the right cursor key t...

Page 33: ...message will be displayed and the voltage will be set to the protection level Press the protection key to set the OVL level Displayed LCD Message PROTECT 1 OVP Use the right cursor key to go to the O...

Page 34: ...ECTION LIMIT message will be displayed and the current will be set to the protection level Press the protection key to set the UCL level Displayed LCD Message PROTECT 1 OVP Use the right cursor key to...

Page 35: ...MIT message will be displayed and the current will be set to the protection level Press the protection key to set the OCL level Displayed LCD Message PROTECT 1 OVP Use the right cursor key to go to th...

Page 36: ...cked Press and hold the protection key lock key to unlock the front panel The lock lamp turns off when the front panel is unlocked Related remote interface commands KEYL OFF ON KEYL Application Checki...

Page 37: ...2C Settings The procedure for setting RS232C is described below Turn on the power switch After turning on power check if the unit is in OUTPUT OFF mode Press the IO Local key for RS232C settings Displ...

Page 38: ...hod Figure 3 3 It is convenient to use an additional adapter cable if the remote device supports only the DB25PIN platform or if you want to use that platform Refer to 1 2 Accessories and Options to c...

Page 39: ...CD Message I O 2 RS485 Press the IO Local Key for the Baud rate settings Displayed LCD Message br 1 9600 bps This can be different depending on the previous state The Baud rate is composed of 9 600 19...

Page 40: ...cket is a port for parallel RS485 wiring Resistance value when using terminating resistors Data Data line 120 1 4W Data Full Up resistance 2K 1 4W Data Full Down resistance 2K 1 4W Figure 3 6 N C Grou...

Page 41: ...RS232C RS485 Wiring Cable Specifications RS485 Pin Diagram Specifications Figure 2 RS232C Pin Diagram Specifications 38...

Page 42: ...splayed LCD Message I O 4 tcp ip Press the IO Local Key for the TCP IP settings Displayed LCD Message 1 Static This can be different depending on the previous state There are two methods for setting a...

Page 43: ...Method Setting Procedure Turn on the power switch After turning on power check if the unit is in OUTPUT OFF mode Press the IO Local Key for the TCP IP settings Displayed LCD Message I O 1 RS232C This...

Page 44: ...n on the power switch After turning on power check if the unit is in OUTPUT OFF mode Press the IO Local Key for the TCP IP settings Displayed LCD Message I O 1 RS232C This can be different depending o...

Page 45: ...tings Displayed LCD Message Response You may receive a response from the device when transmitting setting orders This is unlocked at the time of shipment from the factory The procedure for this settin...

Page 46: ...ent settings can be changed Press the IO Local key once more to disable the voltage and current settings mode Displayed LCD Message output off Limit Display in Output ON mode Turn on the power switch...

Page 47: ...rrent power supply s state in the User Memory Displayed LCD Message STORE NO 01 Use the encoder switch to select the slot you want to use among 01 10 For example turn the encoder switch once clockwise...

Page 48: ...the current power supply Displayed LCD Message RECALL NO 01 Use the encoder switch to select the slot you want to use among 01 10 For example turn the encoder switch once clockwise to save the state...

Page 49: ...ff the output again press the OUTPUT ON OFF key once more Related remote interface commands OUTP OFF ON OUTP ApplicationHow to check the output state and turning it on if it is off OUTP Return value 0...

Page 50: ...D returns to the previous state after a moment Displayed LCD Message NO ERROR The ERR lamp is turned off If there is an error the corresponding error code is displayed Displayed LCD Message ERROR NO 2...

Page 51: ...n can be done through the PC based remote interface Storing on the non volatile permanent memory Calibration data restoration and backup support Easy to control by entering the calibration equipment s...

Page 52: ...ration You can use a general purpose resistor with identical capacity to the electronic load instead Figure 4 1 For Voltage Monitoring During Current Calibration 0 001 0 01 For Ripple Noise Measuring...

Page 53: ...g Resistor shunt Since common current measuring instruments are not suitable for high current capacities current monitoring resistors must be used Select the TCR for 10ppm or below Use a 0 01 high pre...

Page 54: ...4 5 Using Front Panel Calibration Below structure is the procedure of using front panel CALIBRATE Key CALIBRATION KEY Structure 51...

Page 55: ...t Wait until DVM display voltage becomes stable Enters a voltage value that you obtained by reading an external DVM meter If the value is 151 9mV enters as following below LCD Display V LOW 00 15V Thi...

Page 56: ...t calibration connect the power supply output terminal and electronic load and connect the power supply output terminal and shunt lead for current monitoring and then connect shunt s the other lead to...

Page 57: ...cursor key and press the Calibration key LCD Display a high 61 00a Display current can be different Wait until DVM display voltage becomes stable Calculate the current when DVM display voltage became...

Page 58: ...commands as next order if occurs the errors remote calibration is cancelled If error is occurred you must restart remote calibration Voltage Calibraton Send the OFF command to electronic load Send mi...

Page 59: ...mpere After waiteing for minimum 5 minutes more measures DVM s voltage that is connected with current monitoring resistor Send calculated current value that caculated by current value calculating prog...

Page 60: ...supply power off Be able to change to minimum value of voltage current value automatically if you are not use the power supply for long time Be able to lock the power supply s front panel if you are n...

Page 61: ...57...

Page 62: ...el OCP setting maximum value UVL Level 0V OVL Level Limit maximum value UCL Level 0A OCL Level Limit maximum value Output Mode OFF USER MEM CLEAR Operating Power switch on in a pressing FACTORY key Po...

Page 63: ...nt can be output In this case caution is required because it is dangerous LAST STATE Power switch on in a pressing FACTORY key Power switch ON When appearing 1 user mem clear message release FACTORY k...

Page 64: ...NABLE Enable AUTO CURSOR MV mode AUTO CURSOR MV Operating Power switch on in a pressing FACTORY key Power switch ON When appearing 1 user mem clear message release FACTORY key Enter to 3 AUTO CURSOR M...

Page 65: ...TO KEY LOCK mode ENABLE Enable AUTO KEY LOCK mode AUTO KEY LOCK Operating Power switch on in a pressing FACTORY key Power switch ON When appearing 1 user mem clear message release FACTORY key Enter to...

Page 66: ...USE mode OVP USE Operating Power switch on in a pressing FACTORY key Power switch ON When appearing 1 user mem clear message release FACTORY key Enter to 5 OVP USE mode by using cursor key In order t...

Page 67: ...USE mode OCP USE Operating Power switch on in a pressing FACTORY key Power switch ON When appearing 1 user mem clear message release FACTORY key Enter to 6 OCP USE mode by using cursor key In order t...

Page 68: ...nt default value 50Hz 20ms Measures voltage current at every 20ms 100Hz 10ms Measures voltage current at every 10ms 300Hz 3 3ms Measures voltage current at every 3 3ms 1 3KHz 770us Measures voltage cu...

Page 69: ...e more time LCD Display There s no related Remote Interface Command 5 11 CAL BACKUP Recommended calibration interval is 6 monthes and in order to precision of the power supply you must calibrate this...

Page 70: ...TORY key one more time LCD Display There s no related Remote Interface Command 5 13 LOAD DEFAULD This function is to return to default value that shipped the unit from factory You can erase setting an...

Page 71: ...e a value or a code examples MIN MAX for the enclosed parameter A vertical bar separates one of two or more alternative parameters Command finish suffix is LF 0AH and you can change to CR CRLF Maximum...

Page 72: ...FACT AUTO LOC DIS ENA FACT AUTO LOC FACT OVP DIS ENA FACT OVP FACT OCP DIS ENA FACT OCP FACT ADC 5 20 50 100 300 1300 FACT ADC FACT LOAD DEF SYST BEEP SYST ERR SYST VERS SYST TEMP SYST REM 232 l 485 O...

Page 73: ...e voltage value of the output terminals This command also increases or decreases the immediate voltage level using the UP or DOWN parameter by a predetermined amount The command VOLTage STEP sets the...

Page 74: ...ic value ex return value 32 0000 VOLT OVP TRIP This query returns a 1 if the overvoltage protection circuit is tripped and not cleared or a 0 if not tripped Return value 0 not tripped state 1 OVP Trip...

Page 75: ...voltage to above 15V VOLT OVL This query returns the under voltage limit value presently programmed Return value numeric value ex volt ovl return value 15 0000 CURR current UP DOWN This command progra...

Page 76: ...ge ex curr ocp 5 2 Set the OVP Level to 5 2A CURR OCP This query returns the overcurrent protection trip level presently programmed Return value numeric value ex curr ocp return value 5 2000 CURR OCP...

Page 77: ...r current limit value presently programmed Return value numeric value ex curr ucl return value 5 0000 CURR OCL numeric value This command sets the over current limit level numeric value Currently sett...

Page 78: ...alue P 6 5 Measure Measure commands measure the output voltage or current MEASure commands acquire new data before returning the reading Also you don t need DVM Digital Volt Meter and ammeter and you...

Page 79: ...s for current calibration Be careful calibration procedure Refer to 4 6 Remote Interface Calibration current This command save the calibration measuring current value in MIN band and MAX band MIN Cali...

Page 80: ...enable DIS Disable AUTO CURSOR MV mode ENA Enable AUTO CURSOR MV mode ex fact auto cur ena Disable AUTO CURSOR MV mode FACT AUTO CUR Confirm the auto cursor value Return value 0 Disable state 1 Enabl...

Page 81: ...res voltage current at every 50 times 1sec measure precise value 100 Measures voltage current at every 100 times 1sec measure precise value 300 Measures voltage current at every 300 times 1sec measure...

Page 82: ...ror Message Return value e error number message ex syst err return value 222 Out of data 1 About errer refer to 7 Error Messages 2 If you send CLS command every error is cleared But in case of RST com...

Page 83: ...utput state of ON or OFF ON Output on OFF Output off ex1 outp on DC power output on ex2 outp off DC power output off OUTP Confirm the power supply s DC output ON or OFF state Return value 0 Output off...

Page 84: ...ird Products detailed version ex idn return value ODA Technologies PT Series 1 3 1 3 1 2 SN Confirm the power supply serial number This can be applied as a serial number for distribution when developi...

Page 85: ...power supply OVP Maximum OVP value of the power supply OCP Maximum OCP value of the power supply UVL 0V OVL Voltage Limit MAX value UCL 0A OCL Current Limit MAX value Volt Step Minumum value of settin...

Page 86: ...he internal temperature rises more than 70 4 FAN Trip error Occurs when the FAN has stopped 5 AC Input Trip error Occurs when the error of the monitored AC input voltage value is more than 15 6 PFC CA...

Page 87: ...4 6 Remote Interface Calibration for GPIB 20 Ignored min run under volt If you set the max command or value input of calibration without voltage Min calibration operating at first Procedure Min VALUE...

Page 88: ...If you set the max command without Value of current Min calibration operating Procedure Min VALUE MAX VALUE 29 Invalid min value use under curr After sent the command of current Min value and then you...

Page 89: ...It means the error of setting available voltage value 81 Memory limit curr error It means the error of setting available current value 82 Memory max volt error It means the error of the power supply...

Page 90: ...is power supply You may have misspelled the or it may not be a valid command If you are using the short form of the command ex volta 10 The power supply just receive Volt or Voltage You must revise to...

Page 91: ...of this instrument in a wet state could result in electrical shock or fire Avoid vibration or severe impact Make sufficient space at the sides and rear of the power supply for adequate air circulation...