3-6. Programming Under Voltage Limit (UVL)
The encoder switch can change the voltage from 0V up to the maximum value. If the UVL is active,
however, the encoder can change the voltage only from the UVL up to the maximum voltage.
The UVL level values shown below may be different from the actual values.
▌Turn on the power switch.
After turning on power, check if the unit is in "**OUTPUT OFF**" mode.
▌Since the voltage limit must be set higher than the set UVL value before setting
the UVL, the voltage limit must be set first.
▌When setting the voltage, use the left cursor key to move the cursor to the
desired voltage range, and use the encoder switch to select your desired voltage.
The voltage cannot exceed the protection level if the Protection(OVP/UVL/OVL)
settings are active. If you try to exceed the limit in this case, the “PROTECTION
LIMIT” message will be displayed, and the voltage will be set to the protection level.
▌Press the protection key to set the UVL level.
Displayed LCD Message
PROTECT> 1.OVP
▌Use the right cursor key to go to the OVL menu.
Displayed LCD Message
PROTECT> 3.uvl
▌Press the protection key once more in “
protect> 3.UVL
.”
Displayed LCD Message
UVL SET> 00.00v
▌Use the cursor keys to place the cursor on the desired value.
▌Control the encoder switch to change the level value.
The UVL level cannot be set lower than the currently outputted voltage
to prevent operator mistakes.
▌Press the protection key when the change is complete.
Displayed LCD Message
CHANGED
Return to the previous stat after the message above is displayed.
Canceling
▌If you want to cancel the setting, wait for 5 seconds.
Wait for 5 Seconds
Displayed LCD Message
CANCEL…
≫ Related remote interface commands
VOLT:UVL {<voltage>}
VOLT:UVL ?
Application: VOLT:UVL 5
Set UVL Level
Refer to “Initial Setting Value” in “1-6 Check after Power On” for the initial OVP settings
- 29 -