3-9. Programming Over Current Limit (OCL)
The encoder switch can change the current from 0A up to the maximum value. If the OCL is active,
however, the encoder can change the current only from 0A up to the OCL.
The OCL level values shown below may be different from the actual values.
▌Turn on the power switch.
After turning on power, check if the unit is in "**OUTPUT OFF**" mode.
▌Since the current limit must be set lower than the set OCL value before setting
the OCL, the current limit must be set first.
▌When setting the current, use the left cursor key to move the cursor to the
desired current range and use the encoder switch to select your desired current.
The current cannot exceed the protection level if the Protection(OCP/UCL/OCL)
settings are active. If you try to exceed the limit in this case, the “PROTECTION
LIMIT” message will be displayed, and the current will be set to the protection level.
▌Press the protection key to set the OCL level.
Displayed LCD Message
PROTECT> 1.OVP
▌Use the right cursor key to go to the OCL menu.
Displayed LCD Message
PROTECT> 6.Ocl
▌Press the protection key once more in “protect> 6.OCL.”
Displayed LCD Message
OCL SET> 88.00A
▌Use the cursor keys to place the cursor on the desired value.
▌Control the encoder switch to change the level value.
The OCL level cannot be set lower than the currently outputted current
to prevent operator mistakes.
▌Press the protection key when the change is complete.
Displayed LCD Message
CHANGED
Return to the previous stat after the message above is displayed.
Canceling
▌If you want to cancel the setting, wait for 5 seconds.
Wait for 5 Seconds
Displayed LCD Message
CANCEL…
≫ Related remote interface commands
CURR:OCL {<current>}
CURR:OCL?
Application: CURR:OCL 44
Set OCL Level
Refer to “Initial Setting Value” in “1-6 Check after Power On” for the initial OVP settings
Note
- 32 -