▌Enters a voltage value that you obtained by reading an external DVM meter
If the value is 31.607V, enters as following below.
LCD Display
V-high 31.61v
This is an examples for 30V
▌After finishing to enter the value, press the "Calibration" key.
LCD Display
ADC datA f100h
About 500ms, display HEX value instead
of readback value.
LCD Display
2.cal-volt high
Return to the previous state of calibration
▌Volatage calibration is finished. It can be applied next operation of power supply.
Current Calibration
▌Turn off all device and connect all device before calibration
>When current calibration, connect the power supply (+) output terminal and electronic load
and connect the power supply (-) output terminal and shunt lead for current monitoring
and then connect shunt's the other lead to electronic load (-) terminal
>Connect the DVM (+) test lead at the electronic load connction of current monitoring resistor
connect the DVM (-) test lead at the other side of current monitoring resistor
Press hold
▌Power switch on while pressing Calibration key.
▌Power switch ON
Release
▌After self diagnose mode, release the Calibration key
when seeing"1.cal-volt low" Message
▌Using cursor key, pressing Calibration key while "3.cal-curr low" mode.
LCD Display
a-LOW 00.14a
Current Display value can be different.
▌Wait until the voltage displayed in DVM meter be stable.
Wait for 5 minute
- 53 -