Guidance and manufacturer’s declaration –
electromagnetic immunity Table of IEC 60601-1-2:2014
The
INVU by Nuvo™
system is intended for use in the electromagnetic environment specified below. The
customer or the user of the
INVU by Nuvo™
Interface should assure that it is used in such an environment.
IMMUNITY test Test level
Compliance level
Electromagnetic environment
–
guidance
Phase
angle:
1.
0°; 45°;
90°; 135°;
180°; 225°;
270°; 315°
2.
0°
3.
0°
4.
0°
Phase
angle:
1.
0°; 45°;
90°; 135°;
180°; 225°;
270°; 315°
2.
0°
3.
0°
4.
0°
applied to INVU by Nuvo™
during the charging mode only.
Duration:
1. 8.33ms
2. 16.67ms
3. 417ms
4. 5000ms
Duration:
1. 8.33ms
2. 16.67ms
3. 417ms
4. 5000ms
Radiated immunity to power frequency (50/60 Hz) magnetic field
IEC 61000-4-
8:2009
Magnetic
field strength:
30 A/m
Magnetic
field
strength:
30 A/m
Power frequency magnetic
fields should meet the
standards of a typical domicile,
commercial or hospital
environment.
Frequency:
50 HZ &
60 Hz
Frequency 50 HZ &
60 Hz