Appendices
System Specifications
92
Nuvo Group Ltd.
Table 10-3. Electromagnetic Immunity for Non-Life-Supporting Systems
Guidance and manufacturer’s declaration –
electromagnetic immunity Table of IEC 60601-1-2:2014
The
INVU by Nuvo™
system is intended for use in the electromagnetic environment specified below. The
customer or the user of the
INVU by Nuvo™
Interface should assure that it is used in such an environment.
IMMUNITY test Test level
Compliance level
Electromagnetic environment
–
guidance
Immunity to electrostatic discharge (ESD)
IEC 61000-4-2:
2008
± 8 kV contact discharge;
± 2, 4, 8, 15 kV air discharge
± 8 kV contact discharge;
± 15 kV air discharge;
The relative humidity should be
at least 5%. It is recommended
to touch a large metal object
before touching the INVU
device.
Conducted Immunity to electrical fast transients/bursts
IEC 61000-4-
4:2012
Repetition
frequency:
100 kHz
Repetition
frequency:
2 kV
Mains AC power should meet
the standards of a typical
domicile, commercial or
hospital environment. AC
power is applied to INVU by
Nuvo™ during the charging
mode only.
Test voltage: 2 kV
Test voltage: 100 kHz
Line
description:
Line 1 &
Line 2
Line
description:
Line 1 &
Line 2
Conducted immunity to voltage surges
IEC 61000-4-5:
2014
AMD1:2017
Applied
voltage:
±0.5, ±1.0,
±2.0 kV
Applied
voltage:
±0.5, ±1.0,
±2.0 kV
AC power should meet the
standards of a typical domicile,
commercial or hospital
environment. AC power is
applied to INVU by Nuvo™
during the charging mode only.
Surge
application
L1-to-GND;
L2-to-GND
Surge
application:
L1-to-GND;
L2-to-GND
Conducted immunity to voltage dips, short interruptions under minimum 100 VAC, 50 Hz power option
IEC 61000-4-
11:2004
Voltage
reduction,
% of Vnom:
1.
100%
2.
100%
3.
30%
4.
100%
Voltage
reduction,
% of
Vnom:
1.
100%
2.
100%
3.
30%
4.
100%
AC power should meet the
standards of a typical domicile,
commercial or hospital
environment. AC power is
applied to INVU by Nuvo™
during the charging mode only.
Phase
angle:
1.
0°; 45°;
90°; 135°;
180°; 225°;
270°; 315°
2.
0°
3.
0°
4.
0°
Phase
angle:
1.
0°; 45°;
90°; 135°;
180°; 225°;
270°; 315°
2.
0°
3.
0°
4.
0°
Duration:
1. 10ms
2. 20ms
3. 500ms
4. 5000ms
Duration:
1. 10ms
2. 20ms
3. 500ms
4. 5000ms
Conducted immunity to voltage dips, short interruptions under minimum 240 VAC 50 Hz power option
IEC 61000-4-
11:2004
Voltage
reduction,
% of Vnom:
1.
100%
2.
30%
3.
100%
Voltage
reduction,
% of
Vnom:
1.
100%
2.
30%
3.100%
AC power should meet the
standards of a typical domicile,
commercial or hospital
environment. AC power is