Invu by Nuvo™
User Manual
Troubleshooting
NV-UMLS-0012
77
System Errors
Cause
Recommended Actions
abdomen and contact customer
support via the support chat in the
app or at
or
1-800-554-9041 (USA only).
INVU Sensor Band
Battery Level Is Too
Low.
The battery level is too low.
The sensor band battery
level must be at least 20%
to perform a session.
Turn Off the sensor band.
Remove the sensor band from
your abdomen and clean it.
Charge it for at least 30 minutes
using the INVU original charger.
Disconnect the INVU sensor
band from the charger.
Power On the sensor band.
Follow session preparation
instructions and start a new
session.
If the battery level is at least 20%,
you can start a new session
otherwise a message to charge your
sensor band will appear on the
home page.
Please see section 3.3 for charging
instructions.
General Error
An application error
occurred during the
session.
Press the CLOSE button.
Restart the session.
If the problem persists, please
contact customer support via the
support chat in the app or at
or
1-800-
554-9041 (USA only).
Note:
If a problem persists, please contact Nuvo Group customer support via the support
chat in the app or at
or at
1-800-554-9041 (USA only).