Maintenance
Cleaning the INVU Sensor Band™
66
Nuvo Group Ltd.
Power OFF the band by a long press on the power button. (see
“
Powering OFF the INVU
Sensor Band™”
in section 3.2.1).
Remove the band (see
“
Removing the
INVU Sensor Band™”
in section 5.3).
Detach the detachable sensors from their connectors.
Figure 7-4. Detaching the detachable sensors from the
INVU Sensor Band™
Rinse the detached sensors with tap water until no residue of gel is left in the sensor's
cavity.
Let the detached sensors air-dry or use a dry lint-free cloth; then, reattach the sensors
to the band sockets (you should hear an attaching click sound).
Spray all plastic parts with an alcoholic based spray provided in the package and wipe
the spray using a dry lint-free cloth.
Spray the cables with alcoholic based spray provided in the package and wipe the spray
using a dry lint-free cloth.
Spray the outer side of the band with alcoholic based spray provided in the package for
full coverage, and wipe with a dry lint-free cloth.
Note:
We recommend using at least 4 wipes in each step that requires wiping with wipes.
Clean the inner side of the band (the part facing the abdomen), including the Acoustic
sensors and the detachable sensor, with at least 4
CaviWipes™ wipes or equivalent.
Please refer to the chosen cleaning product instructions-of-use, to assure sufficient
cleaning and to avoid infection.
Clean the inner side of the band (the part facing the abdomen), including the Acoustic
sensors and the detachable sensor, with an alcoholic based spray provided in the
package:
➢
One spray for each detachable sensor