Table 10-2. System Compliance Information
–
Enclosure Port Immunity to RF Wireless communications
equipment. Services for which the recommended separation distance is 0.3 m (12 inches)
Test
Frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Immunity
Level
(V/m)
Compliance
Level
(V/m)
385
380
–
390
TETRA 400
Pulse
modulation
b)
18 Hz
27
27
450
430
–
470
GMRS 460,
FRS 460
FM
c)
5 kHz
deviation
1 kHz sine
28
28
710
704
–
787
LTE Band
13, 17
Pulse
modulation
b)
217 Hz
9
9
745
780
810
800
–
960
GSM
800/900
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
b)
18 Hz
28
28
870
930
1 720
1700
–
1990
GSM 1800;
CDMA
1900;
GSM 1900;
DECT;
LTE Band
1, 3, 4, 25;
UMTS
Pulse
modulation
b)
217 Hz
28
28
1 845
1 970
2 450
2400
–
2570
Bluetooth,
WLAN,
802.11
b/g/n
RFID 2450,
LTE
Band 7
Pulse
modulation
b)
217 Hz
28
28
5 240
5
1
00
–
5800
WLAN
802.11 a/n
Pulse
modulation
b)
217 Hz
9
9
5 500
5 785
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave
signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz
may be used because while it does not represent actual modulation, it
would be the worst case.