Know Your INVU Sensor Band™
Standard Operation Duration
14
Nuvo Group Ltd.
Insert the USB type-C plug of the supplied power supply to the Micro USB charging
socket at the side of the plastic controller.
If you connect the charging cable while the device is turned ON, the device will
automatically turn OFF upon charging.
The minimum charging capacity to start a monitoring session is approximately 20%,
which is equivalent to about 45 minutes of monitoring.
The band battery status (charging capacity) is displayed on the INVU
™
App Home screen.
When charging is complete, disconnect the USB type-C plug from the Controller.
Note:
Environmental conditions such as temperature, communication distance,
interference, etc. may affect power capacity and battery consumption. These variations are
typically about ±20%. (see more details in section 9).
Note:
When the LED turns green, it is an indication that the
INVU Sensor Band™
is fully
charged.
Note:
In normal Monitoring Mode,
INVU Sensor Band™
does not have or need any cable
for normal functionality.
Note:
INVU by Nuvo™ wil
l not function if the battery is depleted. If the INVU Sensor
Band™
battery does not charge, contact Nuvo Group customer support.
3.4
Standard Operation Duration
The
INVU Sensor Band™
has a limited usage duration for standard operation.
There is an expiration date mentioned on the
INVU by Nuvo™
Packaging Label, and there is
a limited number of session hours that the
INVU Sensor Band™
can perform.
The
INVU Sensor Band™
can be used until its expiry date is reached or until it reaches its
maximum number of sessions; whichever comes first.
Note:
When the number of sessions of a band reaches or exceeds its maximum hours of
sessions, a notification will appear on the user
’s
application when attempting to initiate a
session. A monitoring session will not start
in this case, and the user should contact Nuvo’s
support.