Chapter 8 Technical Specifications
8.1
Specifications
83
8
J-CY Cytodiagnostic Unit for Nikon Microscopes
Model
J-CY Cytodiagnostic Unit
Optical system
Infinity-corrected CF optical system
Objective: 20X
Variable magnification lens:
2X, 0.5X
Magnification
adjustment system
Quick magnification changeover by rotary solenoid
Operated by C-HS hand switch
Marker section
Quick stamp system (Ink: J-CY refill ink)
Input voltage
12V DC, 1.5A
Power supply
AC adapter
(Power may be supplied directly from the ECLIPSE 55i [through the
cytodiagnostic-unit power supply connector] using the cable set for the 55i &
cytodiagnostic unit.)
Specified AC adapter
Manufacturer:
ILAN ELECTRONICS LTD.
Model: F1650K
Rated input voltage:
100-240V AC, 1.2A max. (when inputting 115V AC),
50/60Hz
Rated output voltage:
12V DC, 3.5A max.
Other:
UL-listed product, GS-approved, CE-certified
AC adapter power cord
•
When used in 100-120V areas:
UL-listed detachable power cord set (3 conductor grounding Type SVT, AWG 18,
3 m long maximum, rated at 125V AC minimum)
•
When used in 220-240V areas:
EU/EN-approved 3-conductor power cord set (3 conductor grounding Type
H05VV-F, 3 m long maximum, rated at 250V AC minimum)
Operating conditions
Temperature: 0
to
40
°
C
Humidity:
85% RH max. (no condensation)
Altitude:
2000 m max.
Degree of pollution:
Degree 2
Installation: Category
II
Indoor use only
Transport/storage
conditions
Temperature: -20
to
60
°
C
Humidity:
90% RH max. (no condensation)
Weight