42
CHECK AND ADJUSTMENT PROCEDURES
:
Polishing Adjustment
8) Change the parameter by the difference between 45.00 mm and the outer diameter.
Change the parameter value with the numeric keypad. Enter the value and press the
button.
Example) When the outer diameter is ø45.15, decrease the parameter value by 0.15.
3
Press the Exit tab to return to the layout screen.
4
Repeat Steps
1
to
3
until the outer diameter becomes ø45.10 ±0.10 mm.