INSTALLATION PROCEDURE
2.1
Installation Location and Instrument Main Body Movement
z
Conditions of installation location
z
Instrument main body movement
CAUTION
• Leave a space of 10 cm or more behind the instrument.
If the cooling fan is blocked, the temperature inside the instrument will rise and it may
cause malfunction.
• Install the instrument on the specified table (optional) or a support which is strong
enough for the weight of the instrument (52 kg).
Vibrations and noises may be produced, and normal processing may not be performed.
WARNING
• Use a hand cart to move the instrument to a different location. At least two persons
are necessary to lift the instrument.
Back injury or stumbling may occur when a single person attempts to move the instru-
ment.
• When lifting the instrument,
do not lift it by its cover. Lift it
by placing hands at front
lower parts (A) and (B), rear
lower parts (C) and (D), and
right and left side lower parts
(E) and (F).
Lifting by the cover may
cause the instrument to
drop resulting in injury or
malfunction.
• Be careful not to get fingers caught when setting the lifted instrument down.
Fingers may get caught between the table and instrument resulting in injury.