27
CHECK AND ADJUSTMENT PROCEDURES
:
Drilling Check and Adjustment
3
3.6.2
Adjusting the hole diameter
■
Adjust the hole diameter in the following procedure when the hole diameter is not ø2.0 mm ±0.1
mm after the drilling confirmation shape is checked.
1
Display the menu screen in service mode.
See “3.2.1 Entering service mode” (page 16).
2
Press the Grinding tab on the menu
screen.
The Grinding screen is displayed.
1) Press the Adjustment button.
2) Press the Hole button.
3
Adjust the hole diameter.
Press the "Diameter adjustment"
numeric field.
Change the parameter value with the
numeric keypad. Enter the value and
press the
button.
When the ø2.0 straight pin does not pass through the hole,
increase the hole diameter.
When the ø2.0 straight pin is loose around the
hole, reduce the hole diameter.
4
Press the Exit tab to return to the layout screen.
5
Repeat processing the lens with drilling confir-
mation shape data and Steps
1
to
4
until the
hole diameter becomes proper.
2
1)
1)
2)
2)
3