11
INSTALLATION PROCEDURE
:
Pipe Arrangement
2
7
Attach a new stocking filter and drain cover to
the tank.
1) Attach the stocking filter to the drain pipe
outlet of the tank.
2) Put the drain cover in the drain pipe out-
let of the tank.
Put the drain cover with the larger side facing
up.
8
Plug the power cord of the pump into the main body.
1) Connect the power cord with label “1” the
extension code with label.
2) Route the extension code through the
hole on the top or rear of the table and
connect it to the outlet for Pump 1 in the
right side of the main body.
3) Connect the power cord with label “2”
with the extension code with label “2”.
4) Route the extension code through the
hole on the top or rear of the table and
connect it to the outlet for Pump 2 in the right side of the main body.
9
Put the tank back into the table.
10
Connect the feedwater hoses to the tank.
1) Push the feedwater hose labeled as
into the connection labeled as Pump 1.
Push the hose until it clicks.
2) Push the feedwater hose labeled as
into the connection labeled as Pump 2.
Push the hose until it clicks.
Stocking filter
Stocking filter
Drain cover
1
Pump
MAX.1A
Pump
2
MAX.1A
1
2