CHECK AND ADJUSTMENT PROCEDURES
:
Drilling Check and Adjustment
z
Adjusting the Drill zero point
When the difference between C and D is 0.1 mm or more, press the "Drill zero point"
numeric field.
Change the parameter value with the numeric keypad. Enter the value and press the
button.
When D > C and the difference between C and D is 1.0 mm,
set the parameter to 0.5.
z
Adjusting the drill hole vertical position
Press the "Drill hole vertical position" numeric field when the L length is not 37.0 mm ±0.1
mm.
Change the parameter value with the numeric keypad. Enter the value and press the
button.
For L = 37.2 mm,
set the parameter to -0.1.
3
Press the Exit tab to return to the layout screen.
4
Repeat processing the lens with drilling confirmation shape data and Steps
1
to
4
until
each value becomes proper.