40
CHECK AND ADJUSTMENT PROCEDURES
:
Polishing Adjustment
3.8.4
Adjusting the polish size
■
Check the polished surface of a lens processed with the circle 45 internal data. If the outer diame-
ter is not within ø45.00 ±0.10 mm, change the Polish size parameter.
1
Check the Polish size parameter.
1) Load the circle 45 internal data and bevel-polish and flat-polish a lens.
2) Measure the outer diameter of the processed lens.
When the outer diameter is not within ø45.00 ±0.10 mm, change the Polish size parameter.
See “3.2.2 Loading the shape for adjustment” (page 17).
See “3.3 Size Check and Adjustment” (page 18).
2
Adjust the polish size.
1) Display the menu screen in service mode.
See “3.2.1 Entering service mode” (page 16).
2) Press the Grinding tab on the menu
screen.
The Grinding screen is displayed.
3) Press the Adjustment button.
4) Press the Size button.
Outer
diameter
Outer
diameter
2)
2)
3)
3)
4)
4)
Summary of Contents for ME 1200