19
CHECK AND ADJUSTMENT PROCEDURES
:
Size Check and Adjustment
3
2
Press the Grinding tab on the menu
screen.
The Grinding screen is displayed.
1) Press the Adjustment button.
2) Press the Size numeric field
The numeric keypad appears.
3) Change the parameter by the differ-
ence from 45.00 mm.
Enter the value with the numeric keypad
and press the
button.
Example)When the outer diameter is ø45.20, reduce the parameter value by 0.20.
3
Press the Exit tab to return to the layout screen.
4
Repeat Steps 1 to 3 until the outer diameter becomes within ø45.00 ±0.05 for beveled
lens and ø45.00 ±0.10 for flat-edged lens.
2
1)
1)
2)
2)
Summary of Contents for ME 1200
Page 6: ...IV ...
Page 10: ...2 REQUIREMENTS Requirements ...
Page 58: ...50 CHECK AND ADJUSTMENT PROCEDURES Clock Check ...