CHECK AND ADJUSTMENT PROCEDURES
:
Surface Measurement (LMU) Calibration
3
5
Press the Surface meas. (LMU) button
to the right of Calibration.
6
The RMU/LMU calibration jig and right
and left surface feelers (LMUs) moves
and calibration starts.
7
When the calibration is complete, a chime sounds and “Calibration is finished” is dis-
played on the information bar. Hold the RMU/LMU calibration jig and press the
button to unlock it for removal.
8
Load the circle 45 internal data and process a lens with the settings as Two Point, Flat,
None (polish), None (S.F.B.), Size 0.00.
Measure the outer diameter of the processed lens to
check the size.
The outer diameter of the flat-edged len must be within
ø45.00 mm
±
0.05.
See “3.3.1 Checking the size” (page 18).
9
Adjust the size when the outer diameter is other than
ø45.00mm ±0.05.
See “3.3.2 Adjusting the size” (page 18).
10
Repeat Steps 7 and 8 until the outer diameter of the lens is within the setting range.