15
3
3.
CHECK AND ADJUSTMENT PROCEDURES
3.1
Check and Adjustment Procedure Flowchart
No
Yes
Yes
Yes
No
No
Load the circle 45 internal data and pro-
cess a lens with the Two Point, Flat, None
(polish), and None (S.F.B.) settings.
Is the lens size
45.00 mm ±0.05 mm?
Is the lens size
45.00 mm ±0.05 mm?
Is the safety-beveled edge
proper
Perform LMU calibration.
Size adjustment
Safety bevel adjustment
Drilling confirmation shape processing
Hole diameter check
Hole depth check
Hole/slot reference, drill zero point, drill hole vertical position check
Front surface offset
Axis check and adjustment
Polish check and adjustment
Bevel check and adjustment
Groove position and depth check and adjustment
Finish
Load the circle 45 internal data and pro-
cess a lens with the Two Point, Flat, None
(polish), and None (S.F.B.) settings.
Load the circle 45 internal data and pro-
cess a lens with the Two Point, Flat, None
(polish), and Medium (S.F.B.) settings.
Summary of Contents for ME 1200