2.4 Installation
CAUTION
• Use the instrument in the following conditions.
- A dust free environment
- A stable place with no tilt
- A vibration and shock free environment
• Install the instrument where it can be used at the following temperature and humidity.
Use conditions
Temperature:
5ºC - 40ºC
Humidity:
Relative humidity does not exceed 50% at a
maximum temperature of 40ºC.
Pressure:
700 hPa - 1060 hPa
• Install the instrument where the operation area specified in p. 2-6 is provided.
• Install the instrument on the specified table (option) or stage which is strong enough for
the weight (42 kg) of the instrument.
Vibrations and noises may be produced, and normal processing may not be performed.
• Use a stage whose size is larger than the measurements of the figure on the right. Also set
the adjuster at the foot of the stage so that the stage can be adjusted without tilt or play.
It is necessary to make openings in the top
to draw the pipe and cord.
Opening A:
For the power cords of the pumps
Opening B:
For the feedwater hoses and drain pipe
• Do not install the instrument on a place exposed to direct sunlight or a place where the
temperature and humidity are high.
• Install the instrument where pollutants such as corrosive gas, acid and salt do not exist.
Corrosion or malfunction of the instrument may occur.
• Be sure to keep 3 cm or more of space in the rear so that the fan will not be blocked.
If the fan is blocked, the temperature of the inside of the instrument will rise, and it may
cause a malfunction.