WF1947/WF1948 Instruction Manual (External Control)
189
6.1.3 Commands Subject to Restrictions (Compatibility with older WF194x
series)
Some commands are restricted when using older WF194x series command.
The restrictions are described below. The command notation is shown in the type 1 command.
For details of each command, see "6.3 Descriptions of Individual Commands."
[Commands recognized as WF1947/WF1948 commands]
When you replace the commands in the type 2 format of older WF194x series with the commands of
WF1947/WF1948, if the behavior differs, replace them with the compatible commands in type 1 format
of older WF194x series. For support status, see the compatibility support items in "6.2 Older WF194x
Series Commands Support List."
[Commands partially restricted to their operations]
(ACP) As the copy of standard waveform, rising sawtooth wave, and falling sawtooth wave, the triangle
wave with the symmetry value set at that time is copied.
(ARB/ARW)
Arbitrary waveform transfer supports only the ASCII mode.
When you transfer data with 1024 or more bytes, ARB/ARW must be prepended to the head.
(ORG) Only the automatic range is supported as the output range setting.
(MRK) Setting of mark wave number Only the setting of auto burst and trigger burst is supported.
(TRE) Only the sweep and burst mode are supported in the status of external trigger input connector for
remote.
(TRG) The gate oscillation mode is not supported.
(RST) Output is made off at the command execution.
(BRO/MDO/SWO)
If the setting is common to both channels, setting the command to change the frequency and amplitude
changes to the setting common to both channels.
(SFC) The sweep function square wave and sine wave sweep are set/queried using the different
implementation method.
[Commands whose parameter specification method must be changed]
(DDV) Half deviation of pulse width modulation must be set.
(ODV) Half deviation of DC offset must be set.
(PDV) The phase deviation setting must be set up to 180.000 deg.
(STO) Memory number must be set/queried within 1 to 10.
(RCL) Memory number must be set/queried within 1 to 10.
(MCO) Memory number must be set/queried within 1 to 10.
(MDL) Memory number must be set/queried within 1 to 10.
Summary of Contents for Wave Factory WF1947
Page 1: ...MULTIFUNCTION GENERATOR Instruction Manual External Control WF1947 WF1948 NF Corporation ...
Page 2: ......
Page 3: ...DA00037649 002 MULTIFUNCTION GENERATOR Instruction Manual External Control WF1947 WF1948 ...
Page 4: ......
Page 20: ...WF1947 WF1948 Instruction Manual External Control xvi ...
Page 21: ...WF1947 WF1948 Instruction Manual External Control 1 1 Preparations before Use ...
Page 28: ...8 WF1947 WF1948 Instruction Manual External Control 8 ...
Page 29: ...9 WF1947 WF1948 Instruction Manual External Control 9 2 Command Descriptions ...
Page 179: ...WF1947 WF1948 Instruction Manual External Control 159 3 Status System ...
Page 189: ...WF1947 WF1948 Instruction Manual External Control 169 ...
Page 196: ...WF1947 WF1948 Instruction Manual External Control 176 4 Error Message ...
Page 199: ...WF1947 WF1948 Instruction Manual External Control 179 ...
Page 200: ...WF1947 WF1948 Instruction Manual External Control 180 5 Specifications ...
Page 202: ...WF1947 WF1948 Instruction Manual External Control 182 5 3 Multi Line Interface Message ...
Page 203: ...WF1947 WF1948 Instruction Manual External Control 183 ...
Page 204: ...WF1947 WF1948 Instruction Manual External Control 184 6 Compatibility of External Control ...
Page 315: ......