WF1947/WF1948 Instruction Manual (External Control)
177
If an error occurs when the external control is performed, the error number is stored in the error queue.
The error numbers and their corresponding messages and error contents are shown in Table 4.1. You can
query the error number and message using “:SYSTem:ERRor?”.
Table 4.1 Error Numbers, Messages, and their Contents
Error
number
Message
Contents (numbers in parentheses are error codes displayed in the
screen)
-102
Syntax error
The received string contains an incorrect syntax.
-108
Parameter not allowed
There are too many parameters.
-109
Missing parameter
One or more parameters are missing.
-110
Command header error
There is an error in the header.
-111
Header separator error
There is an error in the keyword separator of the header.
-113
Undefined header
The received string contains an invalid header.
-120
Numeric data error
There is an error in the numerical parameter.
-130
Suffix error
There is an error in the suffix of the numerical parameter.
-140
Character data error
There is an error in the discrete parameter.
-150
String data error
There is an error in the string parameter.
-160
Block data error
There is an error in the block parameter.
-200 Execution
error
The command cannot be executed.
The relevant error occurs in the
following cases:
• When "*CLS" cannot be executed due to the status of equipment
• When "*RST" cannot be executed due to the status of equipment
• When other commands cannot be executed
-211
Trigger ignored
GET (Group Execute Trigger), "*TRG", and other commands relating
to the trigger are ignored.
The relevant error occurs in the following
cases:
• When the control command is ignored due to the status of equipment
-220
Parameter error
There is an error in the parameter.
The relevant error occurs in the
following cases:
• When there is an error in the prefix/unit of the numerical parameter
• When the specified unit cannot be used due to the status of
equipment
• When the name of arbitrary waveform exceeds 20 characters
• When the name of arbitrary waveform contains the double quotation
(")
• When the number of data points for the arbitrary waveform is
incorrect
When the number of used waveforms exceeds 128 or
total amount exceeds 512 KW
Summary of Contents for Wave Factory WF1947
Page 1: ...MULTIFUNCTION GENERATOR Instruction Manual External Control WF1947 WF1948 NF Corporation ...
Page 2: ......
Page 3: ...DA00037649 002 MULTIFUNCTION GENERATOR Instruction Manual External Control WF1947 WF1948 ...
Page 4: ......
Page 20: ...WF1947 WF1948 Instruction Manual External Control xvi ...
Page 21: ...WF1947 WF1948 Instruction Manual External Control 1 1 Preparations before Use ...
Page 28: ...8 WF1947 WF1948 Instruction Manual External Control 8 ...
Page 29: ...9 WF1947 WF1948 Instruction Manual External Control 9 2 Command Descriptions ...
Page 179: ...WF1947 WF1948 Instruction Manual External Control 159 3 Status System ...
Page 189: ...WF1947 WF1948 Instruction Manual External Control 169 ...
Page 196: ...WF1947 WF1948 Instruction Manual External Control 176 4 Error Message ...
Page 199: ...WF1947 WF1948 Instruction Manual External Control 179 ...
Page 200: ...WF1947 WF1948 Instruction Manual External Control 180 5 Specifications ...
Page 202: ...WF1947 WF1948 Instruction Manual External Control 182 5 3 Multi Line Interface Message ...
Page 203: ...WF1947 WF1948 Instruction Manual External Control 183 ...
Page 204: ...WF1947 WF1948 Instruction Manual External Control 184 6 Compatibility of External Control ...
Page 315: ......