6.3.141 TRD/?TRD
Description
Sets /queries the trigger delay.
Parameters
Trigger delay setting: 0 (0s)
100.00 (100s)
Resolution: 5 digits for more than 1ms, 0.1µs for less than 1ms
Type 2 (For reference)
Settings: [:SOURce]:BM:DELay {<delay>|MINimum|MAXimum}
Query: [:SOURce]:BM:DELay? [MINimum|MAXimum]
Setting example
Sets the trigger delay to 1ms.
Type 1: TRD 1E-3
Type 2 (For reference):BM:DEL 1E-3
Compatibility Information
The minimum value has been changed from the previous WF194x series.(0.3µs
0s)
6.3.142 TRE/?TRE
Description
Selects/queries the status of the external trigger input connector for remote.
•For the Remote state, when the trigger source is EXT and TRE 1,
TRIG/SWEEP IN on the front panel becomes enabled. For local, always enabled.
•At power-on, when the PST(:SYSTem:PRESet) or RST (*RST) command is executed,
it becomes TRE 1.
Parameters
Status selection (0/1)
Type 1 Type 2 (For reference)
0 : OFF (disables TRIG/SWEEP IN)
1 : ON (enables TRIG/SWEEP IN)
Type 2 (For reference)
Settings: :TRIGger[:SEQuence]:ElN:STATe {0|1|OFF|ON}
Query: :TRIGger[:SEQuence]:EIN:STATe?
Setting example
Enables TRIG/SWEEP IN.
Type 1: TRE 1
Type 2 (For reference):TRIG:EIN:STAT ON
Compatibility Information
It is not support oscillation mode except for the sweep and burst.
A trigger source of the SWEEP IN connector signal is not supported.
Summary of Contents for Wave Factory WF1947
Page 1: ...MULTIFUNCTION GENERATOR Instruction Manual External Control WF1947 WF1948 NF Corporation ...
Page 2: ......
Page 3: ...DA00037649 002 MULTIFUNCTION GENERATOR Instruction Manual External Control WF1947 WF1948 ...
Page 4: ......
Page 20: ...WF1947 WF1948 Instruction Manual External Control xvi ...
Page 21: ...WF1947 WF1948 Instruction Manual External Control 1 1 Preparations before Use ...
Page 28: ...8 WF1947 WF1948 Instruction Manual External Control 8 ...
Page 29: ...9 WF1947 WF1948 Instruction Manual External Control 9 2 Command Descriptions ...
Page 179: ...WF1947 WF1948 Instruction Manual External Control 159 3 Status System ...
Page 189: ...WF1947 WF1948 Instruction Manual External Control 169 ...
Page 196: ...WF1947 WF1948 Instruction Manual External Control 176 4 Error Message ...
Page 199: ...WF1947 WF1948 Instruction Manual External Control 179 ...
Page 200: ...WF1947 WF1948 Instruction Manual External Control 180 5 Specifications ...
Page 202: ...WF1947 WF1948 Instruction Manual External Control 182 5 3 Multi Line Interface Message ...
Page 203: ...WF1947 WF1948 Instruction Manual External Control 183 ...
Page 204: ...WF1947 WF1948 Instruction Manual External Control 184 6 Compatibility of External Control ...
Page 315: ......