WF1947/WF1948 Instruction Manual (External Control)
178
Error
number
Message
Contents (numbers in parentheses are error codes displayed in the
screen)
-221
Settings conflict
The correct syntax parameter is received but the command cannot be
executed due to the status of equipment.
The relevant error occurs in the following cases:
•When the sweep function cannot be set in the log because the
equipment is not a frequency sweep
•When arbitrary waveform data in the array format is set in the
arbitrary waveform edit memory
•When the specification of array/control point format of the arbitrary
waveform data is incorrect
• When the specified channel setting cannot be set in 2-channel
equivalence setting due to another channel's restriction (22039)
•When the external modulation/addition input connector cannot be
used for the external modulation because it is used for the external
addition (23129)
•When the external modulation/addition input connector cannot be
used for the external addition because it is used for the external
modulation (23130)
-222
Data out of range
The correct syntax parameter is received but the command cannot be
executed due to the value out of range.
The relevant error occurs in the following cases:
• When a value out of range is set for each register of the status
system
• When 0 is set for the frequency ratio, N/M
• When a value out of range is set for the user-defined unit, m/n
• When the standard waveform of CH2 in the 1-channel machine is
copied to the memory
• When a value out of the setting range is set (23045)
-225
Out of memory
Memory required for execution is insufficient.
The relevant error
occurs in the following cases:
• When arbitrary waveform data cannot be saved due to insufficient
memory
-290
Memory use error
Command cannot be executed because an error occurs relating to
memory.
The relevant error occurs in the following cases:
• When empty memory is specified for the arbitrary waveform number
• When the currently output or used arbitrary wave is deleted
(32004)
-291
Out of memory
Specified memory does not exist.
The relevant error occurs in the
following cases:
• When a value other than 0 to 128 is specified for the arbitrary wave
number
-310
System error
This is a failure of this instrument.
The relevant error occurs in the
following cases:
• When the internal error occurs (24135, 24136, 35005)
-350
Queue overflow
An error occurs but the error is discarded because the error cannot be
entered in the error queue.
-410
Query INTERRUPTED
A new response message entered the message queue, and thus the
previous response message was deleted.
-420
Query UNTERMINATED
The talker was specified, but the message queue has no response
message.
-440 Query
UNTERMINATED
after indefinite response
Query exists after "*IDN?" in the received string ("*IDN?" must be
the last query in the received string).
Summary of Contents for Wave Factory WF1947
Page 1: ...MULTIFUNCTION GENERATOR Instruction Manual External Control WF1947 WF1948 NF Corporation ...
Page 2: ......
Page 3: ...DA00037649 002 MULTIFUNCTION GENERATOR Instruction Manual External Control WF1947 WF1948 ...
Page 4: ......
Page 20: ...WF1947 WF1948 Instruction Manual External Control xvi ...
Page 21: ...WF1947 WF1948 Instruction Manual External Control 1 1 Preparations before Use ...
Page 28: ...8 WF1947 WF1948 Instruction Manual External Control 8 ...
Page 29: ...9 WF1947 WF1948 Instruction Manual External Control 9 2 Command Descriptions ...
Page 179: ...WF1947 WF1948 Instruction Manual External Control 159 3 Status System ...
Page 189: ...WF1947 WF1948 Instruction Manual External Control 169 ...
Page 196: ...WF1947 WF1948 Instruction Manual External Control 176 4 Error Message ...
Page 199: ...WF1947 WF1948 Instruction Manual External Control 179 ...
Page 200: ...WF1947 WF1948 Instruction Manual External Control 180 5 Specifications ...
Page 202: ...WF1947 WF1948 Instruction Manual External Control 182 5 3 Multi Line Interface Message ...
Page 203: ...WF1947 WF1948 Instruction Manual External Control 183 ...
Page 204: ...WF1947 WF1948 Instruction Manual External Control 184 6 Compatibility of External Control ...
Page 315: ......