59
BARGRAPH
Bar Graph Enable
Syntax:
BARGRAPH <state>
Parameters:
The parameter <state> is either 0 or 1. If <state> is 0 then the bargraph is
disabled. If <state> is 1 then the bargraph is enabled.
Function:
This command enables or disables the bargraph feature. If enabled, the
bargraph will be displayed when the units of measurement are not ratio type
of units and the meter is not in INTG mode.
Related Commands:
BARGRAPH?, TONE, TONE?
BARGRAPH?
Bar Graph Enable Query
Syntax:
BARGRAPH?
Parameters:
None
Function:
This query returns a value showing whether the bargraph is enabled or
disabled. Note: The bargraph is not displayed while in either GPIB or RS-232
mode regardless of enabled status.
Returns:
<state>
<state> is of type <number> which represents the integer 0 if the bar graph is
disabled and 1 if the bar graph is enabled.
Related Commands:
BARGRAPH, TONE, TONE?
CALDATE_n?
Calibration Date Query
Syntax:
CALDATE_n?
where n is either A or B
Parameters:
None
Function:
This query returns the calibration date of the detector as read from the
memory module attached to the 2835-C.
Returns:
<model>
<model> is of type <string> in the double quote format.
Related Commands:
ATTNSN_n?, DETMODEL_n?, DETSN_n?
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com